> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/library-preparation/illumina-single-cell/library-preparation-illumina-single-cell-reference_material-list/000009523.md).

# Illumina Single Cell 3’ RNA Prep cDNA QC best practices

**Importance of cDNA QC**

Ensuring the quality of both the cDNA QC product and the final library is critical for optimal sequencing performance in the Illumina Single Cell 3' RNA Prep workflow. Proper QC assessment ensures that the cDNA is of sufficient quality to generate satisfactory final libraries and verifies that upstream steps, including sample preparation and workflow handling, were performed correctly.

**Recommended cDNA QC Best Practices**

When determining the average cDNA fragment size during the "Check QC Product" step of the Illumina Single Cell 3' RNA Prep workflow, perform the following for consistent and accurate assessment across samples and experiments.

**1**. **Timing of QC Assessment**

* Illumina recommends proceeding with library preparation on purified cDNA within 72 hours of completing the "Clean Up cDNA" step for optimal results.
* If cDNA QC cannot be performed within that timeframe, the cDNA QC product generated during the "Clean Up QC Product" step may be stored at -25°C to -15°C for up to four weeks before QC assessment.
* If submitting samples to a core lab or service provider for fragment analysis, submit the cDNA QC Product together with the final libraries. Review the key considerations below for additional information.

**2**. **cDNA Quantification**

Use a Qubit dsDNA High Sensitivity kit to quantify (in ng/µl) 2 µl of each sample.

* **Note**: Do not use the Bioanalyzer or TapeStation to measure concentration.

**3**. **Fragment Analysis**

Use an Agilent Bioanalyzer or TapeStation to determine the average fragment size of the cDNA. Set the region table to 200 bp to 5000 bp for cDNA QC analysis.

If necessary, dilute samples to make sure they are within the recommended concentration range for the trace instrument.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-56b1bd6dad2f74f2db6731e9f12687353c177d66%2Fimage1.png?alt=media)\
![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-667afbb28cade76d8083fb4eb37658d50bc27e70%2Fimage2.png?alt=media)

**Figure 1**. Representative Illumina Single Cell 3' RNA Prep T10 cDNA QC fragment analysis results from a HEK 293T and mouse NIH 3T3 cell line mixture on a Bioanalyzer HS DNA Assay and TapeStation HSD5000 ScreenTape with the appropriate region table assignments from 200 bp to 5000 bp.

* **Note**: These samples were processed from different cell suspensions during different experiments. Variability in peak shape and distribution may occur between instruments and sample types.

#### **Key considerations for proceeding with Illumina Single Cell Library Prep**

* Refer to [Assessing cDNA QC in Illumina Single Cell 3' RNA Prep](https://knowledge.illumina.com/library-preparation/illumina-single-cell/library-preparation-illumina-single-cell-reference_material-list/000009524.md) to determine if cDNA is satisfactory to proceed with Library Prep.
* To ensure high-quality output, **proceed with Library Prep within 72 hours of completing the** "**Clean Up cDNA**" **step**, **even if still awaiting fragment analysis results from the cDNA QC product**.
* Illumina recommends proceeding with Library Prep on purified cDNA as soon as possible after the "Clean Up cDNA" step for optimal results.
* For further guidance on cDNA storage, refer to [Extending cDNA storage time in Illumina Single Cell 3' RNA Prep](https://knowledge.illumina.com/library-preparation/illumina-single-cell/library-preparation-illumina-single-cell-reference_material-list/000009394.md)

**Important note**: Do not use the cDNA QC product as input for Illumina Single Cell Library Prep, as this will result in assay failure. Only the total volume of cDNA recovered from the "Clean Up cDNA" step can be used for library preparation.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                              |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9523), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009523%20-%20Library%20Preparation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/library-preparation/illumina-single-cell/library-preparation-illumina-single-cell-reference_material-list/000009523.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.