> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/library-preparation/illumina-single-cell/library-preparation-illumina-single-cell-reference_material-list/000009517.md).

# The importance of pilot studies for Illumina Single Cell RNA Preparation prior to workflow changes

Illumina Single Cell 3' RNA Prep technology is optimized for efficient and sensitive transcript capture from live cells, and therefore, quick isolation and mild dissociation of cell types is essential and should be confirmed in a pilot experiment before committing to the actual experiment. It is also helpful to complete a pilot prior to scaling up, increasing sample number, or using Supplemental Enrichment and Amplification (SEA) kit applications. Completion of a pilot study before scaling up is especially important when working with any new sample types. Pilot studies are also valuable for labs new to single cell sequencing.

**General Pilot Study Recommendations**

The T2 kits have the same chemistry as the higher cell capture kits, so T2 kits can be used to assess each step of the scRNAseq process and complete any optimization needed, before scaling up to larger reactions that may require a significantly higher read depth and higher output sequencing runs. The following list of recommendations can help with preparation for an effective pilot study before scaling up, when working with a new sample type, or for users new to single cell.

* Purchasing a T2 kit along with the T10, T20, or T100 kits is an economical way to complete a pilot study.
* Processing a small number of samples (1-4 samples) as opposed to starting with large studies with many treatments or conditions.
* Optimizing steps of cell or nuclei preparation for a new sample type including:
  * Dissociation
  * Accurate quantitation of cell/nuclei counts
  * Cell viability assessment in Illumina Single Cell - Cell Suspension Buffer
  * Proper pipetting force
  * Centrifugation speeds
  * Labeling cells
  * Enrichment
  * Use of third party kits to remove debris and/or dead cells
* Comparing wild type (WT) versus treated or mutant conditions
  * Treated/mutant groups may have lower quality mRNA that is inherent to their biology.
  * Discovering this during a small pilot prior to a large-scale study allows for a more informed experiment.
  * It may be necessary to prepare 2+ Illumina Single Cell reactions for the treated/mutant group to achieve the same number of captured cells as the WT.
  * Technical replicates of treated/mutant groups should be sequenced in the same sequencing run and merged using analysis software to be analyzed as a single sample.
* Completing Illumina Single Cell T2 reactions through cDNA QC and Library QC successfully without sending for sequencing can be economical.
  * If taken through sequencing, this will provide even more information to help guide the larger experiment design when scaling up.
  * Review key sequencing metrics to verify data quality is good before deeper sequencing with larger cell numbers including:
    * Passing Cells
    * % mapped
    * % reads in passing cells
    * % mitochondrial reads
    * Duplication rate
    * Median Genes per Cell
    * Median Molecules per Cell
* Alternatively, after completing all sample preparation processes upstream of the Illumina Single Cell workflow, the total RNA can be isolated and run with an Agilent TapeStation RNA Assay to evaluate the quality and to verify the total RNA is not degraded (RIN value ≥ 7 is recommended).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                              |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9517), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009517%20-%20Library%20Preparation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/library-preparation/illumina-single-cell/library-preparation-illumina-single-cell-reference_material-list/000009517.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.