> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/library-preparation/illumina-single-cell/library-preparation-illumina-single-cell-reference_material-list/000009511.md).

# Sample Enrichment Best Practices for Illumina Single Cell 3' RNA Prep

Fluorescence-activated cell or nuclei sorting (FACS or FANS) and magnetic-activated cell sorting (MACS) can be beneficial for the enrichment of target cell populations and can facilitate the exclusion of dead or damaged cells. Enrichment steps also add significant time and can stress cells, causing an overall decrease in cell health. It can be necessary to perform a second live/dead cell sort following cell sorting to ensure sufficient cell viability for the Illumina Single Cell 3' RNA Prep workflow.

**Dead Cell Removal Kits**

It is not recommended to move forward with cell samples that are less than 75% viable. Adding additional cell wash steps is best if only minor viability improvement is needed. If significant improvement is required, it is recommended to either optimize sample preparation processes and reduce sample preparation time, or to incorporate a Dead Cell Removal Kit (eg, Akadeum Life Sciences, Miltenyi Biotec). These kits require starting with a high number of cells (500,000 minimum cells for Akadeum and 1 million cells for Miltenyi).

**FACS and FANS**

It is common for Fetal Bovine Serum (FBS) to be used in pre-sorting buffers and in collection buffers. As an alternative to FBS (which is inhibitory), Bovine Serum Albumin (BSA) can be used in buffers and for tube blocking. It is critical to complete the cell or nuclei wash step according to the user guide after sorting if the sorted samples contain any potentially inhibitory reagents such as FBS.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-8ebd49dc9d5d7b5d14a84e614d8177301598331e%2Fimage1.png?alt=media)

If working with low cell or nuclei numbers and there will be no inhibitory reagents in the final suspension, a low volume of Cell Suspension Buffer or 1X Nuclei Suspension Buffer may be used in the collection tube (200-400 µl). Be sure to include the recommended final concentrations of RNase inhibitor and BSA for the sample type. The 1-2 ml wash step may be eliminated if there are no potential inhibitors. Do a post-enrichment count and determine the count for the entire tube. The cells or nuclei may then be concentrated by centrifugation to pellet the samples, then removing a safe volume of the supernatant so as not to disturb the cell or nuclei pellet. Make sure there is sufficient volume of supernatant remaining in the tube to load into the PIP tube for the kit size being used. Resuspend the samples and determine the new count by adjusting the tube count by the new final volume.

See the following for additional related information.

* [Best practices for cell preparation for Illumina Single Cell prep](https://knowledge.illumina.com/library-preparation/illumina-single-cell/library-preparation-illumina-single-cell-reference_material-list/000009437.md)
* [Best practices for using Illumina Single Cell Nuclei Isolation Kits for Illumina Single Cell kits](https://knowledge.illumina.com/library-preparation/illumina-single-cell/library-preparation-illumina-single-cell-reference_material-list/000009450.md)
* [Best Practices for Nuclei Isolation with alternative nuclei isolation protocols for Illumina Single Cell](https://knowledge.illumina.com/library-preparation/illumina-single-cell/library-preparation-illumina-single-cell-reference_material-list/000009451.md)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                              |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9511), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009511%20-%20Library%20Preparation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/library-preparation/illumina-single-cell/library-preparation-illumina-single-cell-reference_material-list/000009511.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.