> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/library-preparation/illumina-single-cell/library-preparation-illumina-single-cell-reference_material-list/000009473.md).

# Cell input and volume loading recommendations for Illumina Single Cell 3' RNA Prep

**Cell and Nuclei Loading Parameters**

Cell and nuclei loading in Illumina Single Cell 3' RNA Prep kits is optimized to balance high recovery rates and data quality. The table below summarizes optimal loading parameters by kit size.

| Kit Size | Optimal Cell or Nuclei Input\* | Expected Capture Rate | Cell Loading Volume (µl) | RNase Inhibitor Volume (µl) | <p>Target<br>Concentration</p> |
| -------- | ------------------------------ | --------------------- | ------------------------ | --------------------------- | ------------------------------ |
| **T2**   | 5,000                          | ≥ 2,000               | 4                        | 1                           | 1,250 live cells/µl            |
| **T10**  | 17,000                         | ≥ 10,000              | 5                        | 1                           | 3,400 live cells/µl            |
| **T20**  | 40,000                         | ≥ 20,000              | 8                        | 1 - 2                       | 5,000 live cells/µl            |
| **T100** | 200,000                        | ≥ 100,000             | 20                       | 4                           | 10,000 live cells/µl           |

*\**&#x45;xceeding the recommended input increases the risk of a multiplet rate above 8%.

**Notes:**

* Always include RNase inhibitor in the total loading volume.
* Refer to reference guides for loading volumes to ensure optimal capture efficiency.

**Cell and Nuclei Loading Volume** **Considerations**

* Adjust the target concentration if increasing or decreasing input.
* The total loading volume, including RNase inhibitor, must not exceed the specified limit for each kit size. Exceeding recommended volumes may reduce capture efficiency.
* If cells or nuclei cannot be concentrated to the target level, a lower concentration is generally acceptable as long as the total volume remains within recommended limits.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                              |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9473), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009473%20-%20Library%20Preparation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/library-preparation/illumina-single-cell/library-preparation-illumina-single-cell-reference_material-list/000009473.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.