> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/library-preparation/general/library-preparation-general-reference_material-list/000001314.md). Agents can also search and read this documentation through its [MCP server](https://knowledge.illumina.com/~gitbook/mcp).

# Sequences to use for adapter trimming

When performing sequencing on an Illumina instrument, sequences corresponding to the library adapters can be present in the FASTQ files at the 3' end of the reads if the read length is longer than the insert size (see [Adapter trimming - Why are adapter sequences trimmed from only the 3' ends of reads](https://knowledge.illumina.com/software/general-library-prep/software-general-library-prep-reference_material-list/000002905.md) for additional information). To remove these sequences and prevent issues with downstream alignment, adapter trimmingis an option in Illumina FASTQ generation pipelines. Sample sheets generated with [Illumina Experiment Manager](http://support.illumina.com/sequencing/sequencing_software/experiment_manager.html) contain the necessary sequences in the **Settings** section for Illumina kits. Illumina kits in BaseSpace Sequence Hub Run Planning, and Local Run Manager have adapter information built into the software. However, some third-party tools require the adapter sequence for trimming be specified separately. The recommended sequences to use for each Illumina kit are as follows.

**Note:** if only one sequence is listed, that sequence is used for both reads of a paired-end run. Where the sequence differs between Read 1 and Read 2 of a paired end run, the sequence for each read is listed.

TruSeq single index (previously LT) and TruSeq CD index (previously HT)-based kits:

* Read 1: AGATCGGAAGAGCACACGTCTGAACTCCAGTCA
* Read 2: AGATCGGAAGAGCGTCGTGTAGGGAAAGAGTGT

AmpliSeq for Illumina; Illumina DNA Prep (M) Tagmentation (previously Nextera DNA Flex); Illumina DNA Prep with Enrichment (S) Tagmentation (previously Nextera Flex for Enrichment); Nextera XT; TruPath Genome; TruSight Enrichment; Illumina Rapid RNA kits; Illumina RNA Prep with Enrichment, Ligation:

* CTGTCTCTTATACACATCT

Illumina DNA PCR-Free Prep, Tagmentation

* CTGTCTCTTATACACATCT+ATGTGTATAAGAGACA

Illumina Stranded mRNA Prep, Ligation and Illumina Stranded Total RNA Prep Ligation with Ribo-Zero Plus:

* ACTGTCTCTTATACACATCT

TruSeq Small RNA:

* TGGAATTCTCGGGTGCCAAGG

**Notes:**

* The full adapter sequences for various Illumina library preparation kits can be found in the [Illumina Adapter Sequences Document](https://support.illumina.com/downloads/illumina-adapter-sequences-document-1000000002694.html).
* Illumina Stranded mRNA Prep, Ligation and Illumina Stranded Total RNA Prep Ligation with Ribo-Zero Plus require additional trimming of a T overhang at the 5’ end of the library. See [Trimming T-overhang options for Illumina Stranded mRNA and Illumina Stranded Total RNA workflows](https://knowledge.illumina.com/software/general/software-general-reference_material-list/000003227.md) for further details.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                              |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1314), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001314%20-%20Library%20Preparation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/library-preparation/general/library-preparation-general-reference_material-list/000001314.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.