> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/library-preparation/general/library-preparation-general-reference_material-list/000001146.md).

# Best practices for storage of bead normalized Nextera XT libraries

The Nextera XT protocol includes an optional bead-based normalization step for libraries after the PCR clean-up step in the protocol.

Because the bead-based normalized libraries are single-stranded after elution from the normalization beads, the libraries are not as stable as double-stranded DNA libraries, are more susceptible to degradation, and can adhere to the walls of tubes.

After bead-based normalization, **libraries can be stored for up to 1 week at -25°C to -15°C**.

To make sure that all library strands are completely denatured, **perform a heat denaturation of the libraries immediately before sequencing**, as follows, with the diluted library pool just prior to loading into the sequencing cartridge.

1. Incubate the denatured and diluted library at 96°C for 2 minutes (using a pre-heated heat block).
2. After the heat incubation, invert the tube 1-2 times to mix.
3. Quickly move the library to an ice water bath for 5 minutes. The quick cooling step helps to lock the sample in its single-stranded form.
4. Proceed immediately to cluster generation.

**Note**: For longer term storage of libraries, stop immediately after the PCR clean-up step when the libraries are still double-stranded, and store the SGP plate at -25°C to -15°C. Aliquot as needed to avoid multiple freeze-thaw cycles.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                              |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1146), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001146%20-%20Library%20Preparation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/library-preparation/general/library-preparation-general-reference_material-list/000001146.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.