> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/library-preparation/general/library-preparation-general-faq-list/000010168.md).

# Workflow and Sequencing recommendations FAQ for TruPath Genome

**Note**: Trupath Genome is compatible with the NovaSeq X Series instruments only, with a minimum control software version of 1.4.

**What is the mechanism by which TruPath Genome libraries are made?**

Library preparation uses flow cell-bound transposomes which capture and tagment long molecules of dsDNA as they are flowed onto the flow cell surface. Clustering and SBS are as standard, resulting in high-quality short-read sequencing.

**What changes to the sequencing workflow are required for TruPath?**

Flow cell prep mix (FP1) is added to CP1 position and Flow Cell Transposome (FT2) is added to CP2 position. See the [Product Documentation](https://support.illumina.com/downloads/trupath-genome-product-documentation-200065852.html) for full details.

**Can FP1, FT2, and FTB reagents be refrozen after thawing?**

If running on the same day, keep these reagents at 4°C for up to 4 hours. Otherwise, remove these reagents from the cartridge, and store in capped tubes at -20°C until subsequent use. Up to 5 freeze-thaws are supported.

**How many samples are supported per sequencing run?**

2 samples per C2 flow cell and 8 samples per C8 flow cell (1 sample per lane for both flow cells). Therefore, double per run with 2 flow cells on a NovaSeq X Plus; 16 samples per C8 run and 4 samples per C2 run.

**Can TruPath Genome runs utilize NovaSeq X v1.4 staggered start?**

TruPath Genome runs can be staggered with another TruPath or any other regular flow cell on the NovaSeq X.

**What is the recommended read length?**

The recommended read length is 2x151 bp. Run setup in BaseSpace Run Planning requires 2x151 run set up with no index reads (use "0" for index1 and index2).

**Are indexes sequenced**\*\*?\*\*

TruPath does not require indexing, as only one sample per flow cell lane is supported.

**Is PhiX required? What happens if it is added?**

No. PhiX DNA or PhiX library (eg, Illumina PhiX control v3, cat # FC-101-3001) must **not** be added to the human gDNA samples.

**Is a custom recipe required?**

C2 and C8 flow cells have an RFID to load the correct recipe; this will require the V1.4 software installed on the NovaSeq X to work.

A custom recipe cannot be loaded when running a C2 or C8 flow cell.

**Does the above require selecting custom sequencing primers? ie, should the check boxes for custom primer use be selected during run setup?**

No, do **not** select the check boxes (they must be kept blank) for Custom Primers during run setup. The flow cell RFID will load the correct sequencing recipe for TruPath Genome libraries, which instructs the instrument to use the contents of the CP1 and CP2 wells at the appropriate times. The instrument software will give an error if custom primers, custom recipes, or indexes are selected for runs using TruPath flow cells.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-b0f23208c3386cd2348b5c5288c4572afd467cfe%2Fimage1.png?alt=media)

**What is the total sequencing time?**

Both C2 and C8 TruPath Genome sequencing runs will complete within 29 hours.

**Is there anything specific required for the BCL Convert settings for TruPath sequencing ›runs?**

* **Read lengths must be**: 151, 0, 0, 151
* **Library prep kit:** choose one:
  * Illumina TruPath Genome Prep (C2, 2 samples)
  * Illumina TruPath Genome Prep (C8, 8 samples)
* **Index Adapter kit:** Not Applicable
* For physical run setup, only sequence 1 DNA sample per lane.

**What is the phenotype if there is a swap of the reagents within the CP1 and CP2 positions?**

The sequencing run will fail. If the swap is detected prior to loading the cartridge onto the sequencer, reagents can be removed and placed back into their respective tubes. CP1 and CP2 wells must then be washed three times with 4 ml water, before adding the reagents back to the correct positions.

**What are the sequencing expectations for TruPath Genomes runs?**

The expectations for TruPath Genome sequencing runs are as follows:

| Flow cell | %Q30 | Yield/Lane | Yield/Total |
| --------- | ---- | ---------- | ----------- |
| C8        | 85%  | 241 Gb     | 1928 Gb     |
| C2        | 85%  | 241 Gb     | 482 Gb      |

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                               |
| :---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #10168), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000010168%20-%20Library%20Preparation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/library-preparation/general/library-preparation-general-faq-list/000010168.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.