> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/library-preparation/general/library-preparation-general-faq-list/000009943.md).

# Control (GDMC and SFMC) FAQ for the Illumina 5-Base Whole Genome Sequencing and Enrichment kits

**What controls are needed in the prep? What is the GDMC** (**Genomic DNA Methylation Control**) **and SFMC** (**Short Fragment Methylation Control**)**? Why are they mixed with the sample? How do they work in the assay?**

* GDMC is a combination of pUC19 (a fully methylated genome) and Lambda (a fully unmethylated genome). It is mixed with the sample prior to shearing to help judge the conversion success of the sample.
* SFMC is the same pUC19 and Lamba genomes; however, it is pre-sheared to mimic cfDNA and is spiked into cfDNA samples prior to starting the workflow.

**Is it correct that the only difference between SFMC and GDMC is the fragment size** (**ie, SFMC is sheared to match cfDNA fragment size and gDNA is high molecular weight control DNA**)**?**

* This is correct.

**Are the controls added to each sample** (**eg, to all 96 samples on a plate**) **or are they a separate sample** (**eg, 2 of 96 samples on the plate are controls?**) **Does this change for the Enrichment workflow?**

* Each gDNA sample is combined with 2.5 µl of GDMC (Genomic DNA Methylation Control).
* Each cfDNA sample is combined with 2.5 µl SFMC (Short Fragment Methylation Control).
* The spike-in controls will be spiked in with the DNA samples and are intended to be library-prepared and sequenced along with the library. The customer will not need to sequence them separately.
* This is the same for both the WGS and the Enrichment kit workflows; however, for the Enrichment workflow, to detect GDMC or SFMC, the Methylation Control Panel of oligos must be also purchased and used in the hybridization. For more details, see [Methylation Control Panel FAQ for the Illumina 5-Base WGS and Enrichment kits](/library-preparation/epigenetics-library-prep/library-preparation-epigenetics-library-prep-faq-list/000009938.md).

**Is an unconverted version of each sample needed in addition to the enzyme**-**converted sample for analysis purposes?**

* No, this is not required for the analysis.

**Is it known how alternative controls would perform?**

* Alternate controls are not recommended. The current DRAGEN software is designed to use the Illumina-provided controls only.

**Is pUC19 methylated at only CpGs Cs or all Cs?**

* Only CpG Cs are methylated. Control genomes also do not have any hmC.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                              |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9943), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009943%20-%20Library%20Preparation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter, and the optional `goal` query parameter:

```
GET https://knowledge.illumina.com/library-preparation/general/library-preparation-general-faq-list/000009943.md?ask=<question>&goal=<endgoal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is optional and describes the broader end goal you are ultimately trying to accomplish on behalf of the user. GitBook uses it to tailor the answer towards what is most useful for that goal.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.
