> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/library-preparation/general/library-preparation-general-faq-list/000009940.md).

# Consumable FAQ for the Illumina 5-Base Whole Genome Sequencing (WGS) and Enrichment kits

**What are the kit configurations for the Whole Genome Sequencing (WGS) and Enrichment kits?**

**WGS**

* 24 sample kit
* Indexes required, but sold separately

**Enrichment**

* 24 sample enrichment kit
* Indexes required, but sold separately

**What are the key/uncommon requirements for user-supplied consumables and equipment?**

**WGS or Enrichment kits**:

* \[gDNA] Covaris LE220 Plus, or other sonication equipment + tubes
* \[cfDNA] Agilent Fragment Analyzer or Agilent TapeStation + reagents for input quantification
* Hard-shell PCR plates (250 ul min volume)
* DynaMag Magnet (+/- skirted)
* Qubit + Qubit dsDNA Broad Range kit
* High-Speed Thermal Mixer, or equivalent instrument that can shake up to 2200 rpm

**Notes**:

* No additional equipment or consumables are required for the Enrichment workflow compared to the WGS kit, outside of purchasing an enrichment panel of oligos. Hybridization is performed on a thermal cycler.
* Consult the library prep Product Documentation for a full list of required equipment and consumables.

**Which enzyme is the IME conversion enzyme?**

* The details of the enzyme are proprietary to Illumina.

**Can strip tubes be used instead of 96 well plates?**

* No, Illumina does not recommend using strip tubes. They do not have the same thermal conductance as 96 well plates, which could make thermal cycling less robust. This could lead to poor methylation conversion, for instance. In addition, all of the shaking steps have been optimized for plates, not tubes.

**What is the supported number of freeze thaws of Illumina reagents?**

* Three freeze/thaws.

**Are any reagents very sensitive in this protocol? Eg, to freeze-thaw, precipitates, etc.?**

* The Illumina Methylation Enzyme (IME, the conversion enzyme) is sensitive to ethanol contamination and bead carryover. It is critical to make sure there is no residual ethanol or beads in the sample before the conversion step.
* IME can be staged on ice for >30 minutes when running the protocol. But it should be stored long term at the recommended temp of -20°C, otherwise its conversion activity will be reduced.

**Can the volume of the MRR reagent used be modified?**

* No, modifications to the MRR volume can have negative effect on methylation conversion and therefore should not be altered.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                              |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9940), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009940%20-%20Library%20Preparation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/library-preparation/general/library-preparation-general-faq-list/000009940.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.