> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/library-preparation/general/library-preparation-general-faq-list/000009598.md).

# How many samples can be prepared with the Illumina Exome 2.5 panel?

The Illumina Exome 2.5 Panel is used with the [Illumina DNA Prep with Exome 2.5 Enrichment](https://www.illumina.com/products/by-type/sequencing-kits/library-prep-kits/dna-prep-exome-enrichment.html), [Illumina FFPE DNA Prep with Exome 2.5 Enrichment](https://www.illumina.com/products/by-type/sequencing-kits/library-prep-kits/ffpe-dna-prep-exome-enrichment.html), and [Illumina Rapid RNA Prep with Exome 2.5 Enrichment](https://www.illumina.com/products/by-type/sequencing-kits/library-prep-kits/rapid-rna-prep-exome-enrichment.html) library preparation kits.

**Illumina DNA Prep with Exome 2.5 Enrichment**

The Twist Bioscience for Illumina Exome 2.5 panel oligos tube contain enough material for 96 samples, including overfill, for automation. If preparing samples manually, there will be enough panel volume for additional hybridization reactions if desired. In the manual workflow, 96 samples requires 8 hybridization reactions at 4 µl each = 32 µl panel needed for 96 samples. The guaranteed fill volume is 68 µl.

Note that this library preparation kit only supports 12-plex enrichments, with each enrichment reaction requiring 4 µl of probe panel. One tube of the Illumina Exome 2.5 panel is provided per order of Illumina DNA Prep with Exome 2.5 Enrichment, (S) Tagmentation Set B (96 Samples, 12-plex) (catalog 20077595), or per order of Illumina DNA Prep with Exome 2.5 Enrichment, (S) Tagmentation Set D (96 Samples, 12-plex) (catalog 20077596). Index adapters plus UMI anchors are also included.

**Illumina FFPE DNA Prep with Exome 2.5 Enrichment**

For samples prepared with automation, 4 tubes of probes are required when preparing 192 samples to for sufficient overfill. Library preparation and enrichment reagents for 192 samples for automation can be purchased by ordering: Illumina Cell-Free and FFPE DNA Prep with Enrichment, Ligation (192 Samples, 4-plex), Cloud Analysis (catalog 20104103) or Illumina Cell-Free and FFPE DNA Prep with Enrichment, Ligation (192 Samples, 4-plex), On-premises (catalog 20104104). Index adapters plus UMI anchors and the probe panel are sold separately.

For samples prepared manually, each Illumina Exome 2.5 panel tube contains a volume of 68 µl, which is guaranteed to support 20 hybridization reactions. Each hybridization reaction will contain 4 samples (4-plex) and require 3 µl of probes. Therefore, 1 tube of probes can support 80 samples with the manual workflow.

**Illumina Rapid RNA Prep with Exome 2.5 Enrichment**

This bundled kit supports the library prep and enrichment of 96 samples at 3-plex enrichments (32 enrichment reactions). Each kit order provides 1 tube of the Twist Bioscience for Illumina Exome 2.5 panel. 1 µl of probe is used per enrichment reaction, therefore 32 µl are used in total of the 68 µl provided, when processing 96 samples.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                              |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9598), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009598%20-%20Library%20Preparation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/library-preparation/general/library-preparation-general-faq-list/000009598.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.