> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/library-preparation/general/library-preparation-general-faq-list/000003467.md).

# Single or double-sided bead clean ups with Illumina DNA PCR-Free

**Why are two rounds of IPB clean ups performed for each sample?**\
This protocol uses two rounds for a double-sided bead clean up. The first round removes the larger than desired DNA fragments from the library, while the second round removes the smaller fragments from the library. Narrow library insert size allows for more consistent clustering and leads to high-quality sequencing metrics.

**Why is a two-sided size selection used for the standard workflow? Why not keep higher fragment sizes?**\
Read 2 Q30 scores and other performance metrics can be negatively impacted by larger fragments, so these are removed in the standard library prep workflow.

**Can a single-sided 1.8x (or more) bead clean up be used instead of the two-sided clean up?**

* A single-sided clean up leads to a change in the fragment size distribution of the final library. Though this is necessary for the low input protocol, some metrics (such as Q30 and insert size) may be reduced. Outside of the low input protocol, Illumina does not recommend performing a single-sided clean up under normal testing conditions.
* For more details, see the app note [Tunable insert sizes with Illumina DNA PCR-Free Prep, Tagmentation](https://www.illumina.com/content/dam/illumina-marketing/documents/products/appnotes/illumina-dna-pcr-free-insert-size-app-note-770-2020-004.pdf).

**Why is the size selection different for low DNA input (25 ng-99 ng) and higher DNA input (100 ng-2000 ng)?**

Optimal size selection delivers consistent library fragments and the best quality. If no modification was made, at the lowest inputs, the workflow would generate very low yield and significantly smaller fragment sizes. To improve yield and fragment size at low inputs, the single-sided clean up was implemented. This change has minimal negative impact on quality and ensures enough DNA can be obtained.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                              |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #3467), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000003467%20-%20Library%20Preparation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/library-preparation/general/library-preparation-general-faq-list/000003467.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.