> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/library-preparation/general-library-prep/library-preparation-general-library-prep-reference_material-list/000001545.md).

# What is the PhiX Control v3 Library and what is its function in Illumina Next Generation Sequencing

The PhiX Control v3 Library (commonly referred to as PhiX, [FC-110-3001](http://www.illumina.com/products/phix_control_v3.html)) is derived from the small, well-characterized bacteriophage genome, PhiX. It is a concentrated Illumina library (10 nM in 10 µl) that has an average size of 500 bp and consists of balanced base composition at \~45% GC and \~55% AT.

* **Note**: this is distinct from the PhiX Indexed control; see [PhiX Indexed Control (1000 Cycle) Product Information](https://github.com/illumina-swi/illumina-knowledge/tree/master/articles/Knowledge/PhiX-Long-Read-Formulation/README.md) for more information.

**Calibration Control**

The PhiX Control v3 Library can be sequenced alone and serves as a calibration control for:

* Cross talk matrix generation
* Phasing and prephasing calculations
* Examining the overall performance of Illumina sequencing platforms

For detailed information, see:

* [PhiX loading concentrations for validation runs on Illumina sequencing platforms](/instrumentation/general/instrumentation-general-reference_material-list/000001536.md)
* [How to set up a PhiX validation run with Local Run Manager](/instrumentation/general/instrumentation-general-reference_material-list/000001318.md)
* [How to set up the sample sheet for a PhiX validation run on the MiSeq using IEM](/instrumentation/miseq/instrumentation-miseq-reference_material-list/000001244.md)
* [How to set up a PhiX validation run on MiSeq i100 Series](https://github.com/illumina-swi/illumina-knowledge/tree/master/articles/Knowledge/How-to-set-up-a-PhiX-validation-run-on-MiSeq-i100-Series/README.md)
* [How to set up a PhiX validation run on the NextSeq 500/550 in manual mode](/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-reference_material-list/000001532.md)
* [How to set up a PhiX validation run on the NextSeq 1000/2000 sequencing platform](https://github.com/illumina-swi/illumina-knowledge/tree/master/articles/Knowledge/How-to-set-up-a-PhiX-validation-run-on-the-NextSeqTM-1000-2000-sequencing-platform/README.md)
* [How to set up a PhiX validation run on the NovaSeq 6000 sequencing platform](https://github.com/illumina-swi/illumina-knowledge/tree/master/articles/Knowledge/How-to-set-up-a-PhiX-validation-run-on-the-NovaSeq-6000-sequencing-platform-support-bulletin/README.md)
* [How to set up a PhiX run on the NovaSeq X and NovaSeq X Plus sequencing platform](https://support-docs.illumina.com/IN/NovaSeqX_DnD-PhiX/Content/NovaSeqX/PhiX-NVX.htm)

**Run Quality Monitor**

Due to its balanced nucleotide composition, the PhiX Control v3 Library is also an ideal sequencing control (typically with ≥ 1% spike-in) for run quality monitoring; eg, cluster generation, sequencing, and alignment.

**Color Balancing**

For low diversity libraries, the PhiX Control v3 Library provides balanced fluorescent signals at each cycle to improve the overall run quality. For detailed information, see [What is nucleotide diversity and why is it important?](/instrumentation/general/instrumentation-general-reference_material-list/000001543.md)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                              |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1545), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001543%20-%20Library%20Preparation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter, and the optional `goal` query parameter:

```
GET https://knowledge.illumina.com/library-preparation/general-library-prep/library-preparation-general-library-prep-reference_material-list/000001545.md?ask=<question>&goal=<endgoal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is optional and describes the broader end goal you are ultimately trying to accomplish on behalf of the user. GitBook uses it to tailor the answer towards what is most useful for that goal.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.