> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/library-preparation/general-library-prep/library-preparation-general-library-prep-faq-list/000009000.md).

# How to prepare < 2 nM libraries for sequencing on NovaSeq X/X Plus

The tables in the [Denature and Dilute Protocol Generator for NovaSeq X Plus](https://support-docs.illumina.com/IN/NovaSeqX_DnD-A/Content/NovaSeqX/protocol-A-NVX.htm) provide the µl of 2 nM libraries + µl of RSB to combine to obtain the specified Final Loading Concentration (pM) after the addition of NaOH and Pre-load Buffer. The following math accounts for the later additional dilutions with NaOH and Pre-Load Buffer. Note that sequencing of libraries less than 2 nM has not been validated by Illumina and is not supported.

1. Measure the library pool concentration in pM (Y)
   * eg, 1000 pM
2. Determine the final target loading concentration in pM (X)
   * eg, 180 pM
3. Calculate the volume of library and RSB required using the following formula:
   1. **(Y pM) (Z µl) = (5\*X) (volume needed by flow cell type)**
      * Volume needed by flow cell type = **34 µl** for 10B or 1.5B, and **56 µl** for 25B flow cells
   2. Volume library needed = Z
   3. Volume RSB needed = (Volume needed by flow cell type) - Z
4. Move to **step 4** "**\[Optional] Spike in 1-2% nondenatured PhiX as follows**" [of the "Dilute Libraries and Add PhiX Control" section of the protocol](https://support-docs.illumina.com/IN/NovaSeqX_DnD-A/Content/NovaSeqX/protocol-A-NVX.htm) if adding PhiX, or the "**Denature Libraries**" section if not adding PhiX.

Example calculation for 1.5 B or 10B flow cell:\
(1000 pM) (Z µl) = (5 \* 180 pM) (34 µl)

Volume library needed (Z) = **30.6 µl**\
Volume RSB needed = 34 - 30.6 = **3.4 µl**

Example calculation for 25B flow cell:\
(1000 pM) (Z µl) = (5 \* 180 pM) (56 µl)

Volume library needed (Z) = **50.4 µl**\
Volume RSB needed = 56 - 50.4 = **5.6 µl**

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                              |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9000), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009000%20-%20Library%20Preparation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/library-preparation/general-library-prep/library-preparation-general-library-prep-faq-list/000009000.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.