> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/library-preparation/epigenetics-library-prep/library-preparation-epigenetics-library-prep-faq-list/000009945.md).

# Input recommendations for the Illumina 5-Base WGS and Enrichment kits

**What is the cfDNA and gDNA amount range for the library prep**? **Any low or high limit**?

* cfDNA: 1 - 20 ng recommended
* gDNA: 50 - 100 ng recommended
* To clarify: the gDNA assay is designed to **start** with 50-100 ng of material going into sonication. gDNA that is sheared should have an average size of \~450 bp.

**What is the input amount required for the targeted sequencing/enrichment workflow**?

* The DNA input amount is the same as the non-enrichment workflow. The main difference is the additional PCR cycles.

**What happens if more than the recommended input amount is used**?

* Too much input saturates the enzyme and results in under-conversion.

**Is it possible to run low input gDNA samples**, < **50 ng**?

* The Assay is optimized for 50 - 100 ng inputs. Lower inputs may reduce library complexity and thus impact single nucleotide variant (SNV) and Indel recall and precision. Illumina can recommend the following as a starting point for optimization. Labs will need to optimize and validate the use of low input in this assay:

<table data-header-hidden><thead><tr><th valign="top"></th><th valign="top"></th></tr></thead><tbody><tr><td valign="top">Section</td><td valign="top">Modification</td></tr><tr><td valign="top">Prepare Sheared gDNA</td><td valign="top">Dilute GDMC 1:10 before adding 2.5 µl to gDNA for shearing</td></tr><tr><td valign="top">Ligate Adapters</td><td valign="top">Add 2 µl of UMI3 adapters instead of 5 µl</td></tr><tr><td valign="top">Clean Up Ligation</td><td valign="top">Perform a second consecutive 0.8x bead clean up after ligation</td></tr><tr><td valign="top">Index PCR</td><td valign="top">Run PCR for 9 cycles, instead of 6 cycles</td></tr><tr><td valign="top">Clean Up Index PCR</td><td valign="top">Add 70 µl of IPB instead of 90 µl</td></tr></tbody></table>

* The diluent for GDMC is RSB which is provided in ample excess volume in the kit.

**How should the input be quantified**?

* **gDNA**: Illumina generally recommends fluorometric methods of quantification, such as Qubit dsDNA Broad Range Assay Kit for gDNA input quantification (catalog# Q32851 or Q32854). gDNA concentration can also be assessed by TapeStation using the Genomic DNA ScreenTape, although gDNA quality information is not required.
* **cfDNA**: quantify only the mononucleosomal peak (\~ 75 bp to 250 bp) using a size-based quantification method such as the Agilent Fragment Analyzer or Agilent TapeStation. Details are in the library prep Product Documentation, "DNA Input Recommendations" section.

**Which samples types are expected to be compatible with the Illumina 5-base solution**?

* The assay is compatible with multiple sample types, including cell-free DNA, and DNA extracted from blood, cell lines, or fresh-frozen tissue.

**Is it possible to start with degraded or FFPE DNA input and skip/reduce fragmentation**\*\*?\*\*

* Illumina has demonstrated success of FFPE with Illumina 5-Base in [this flyer.](https://www.illumina.com/content/dam/illumina/gcs/assembled-assets/marketing-literature/illumina-5-base-dna-ffpe-flyer-m-gl-04137/ilmn-5-base-dna-prep-ffpe-aacr-flyer-m-gl-04137.pdf)
* An application note on FFPE is found [here](https://www.illumina.com/content/dam/illumina/gcs/assembled-assets/marketing-literature/illumina-5-base-ffpe-app-note-m-gl-04497/illumina-5-base-dna-ffpe-app-note-m-gl-04497.pdf).
* An updated protocol including full support of FFPE input for the WGS workflow is available in document # 200073497 **v01**.
* Note that FFPE is **not** supported for the Enrichment workflow.

**Which species of input are supported**?

* All species are compatible with the Illumina 5-Base DNA Prep kit.
* For the Illumina 5-Base DNA Prep with Enrichment kit, only human (hg19, hg38) and mouse (mm10, mm39) custom panel designs are supported in DesignStudio. For other species or other features not available through DesignStudio, please contact your Illumina Sales Representative to connect you with Illumina’s Concierge Design team.

**Which blood collection tubes are supported for cfDNA input**?

* Apostle MiniMax Cell-Free DNA Blood Collection Tube (BCT)
* Qiagen PAXgene Blood ccfDNA Tubes
* Streck Cell-Free DNA BCT
* BD Vacutainer Plastic Blood Collection Tubes with K2EDTA

**Does Illumina have recommendations for gDNA or cfDNA isolation kits**?

**gDNA**:

Illumina has tested but not fully validated the following kits; each lab must validate performance in their lab. Alternative extraction kits are also acceptable. Illumina recommends DNA extraction methods leveraging silica membranes for purification.

* **Whole Blood**: QIAwave DNA Blood & Tissue Kit (Qiagen, Hilden, Germany; part number 69554)
* **Fresh frozen tissue**: QIAamp Fast DNA Tissue Kit (Qiagen, Hilden, Germany; part number 51404)
* **Cell-line DNA**: NucleoSpin Tissue DNA Kit (Machery-Nagel, Duren, Germany; part number 740952.50)

**cfDNA**:

Illumina has validated the following kits:

* QIAGEN QIAamp Circulating Nucleic Acid Kit (without carrier RNA)
* QIAGEN QIAamp MinElute ccfDNA Midi Kit
* Zymo MAGicBead cfDNA Isolation Kit

**Are there any known inhibitors of the prep?**

* Ethanol is a known inhibitor of the conversion step. That is why the workflow has steps to prevent ethanol carryover into the conversion step.
* Illumina does not recommend phenol chloroform extraction, as chemical carryover can negatively impact performance. The phenotype of this type of carryover is that the methylation conversion rates are highly variable from sample-to-sample with no apparent pattern.
* K2EDTA tubes have been validated and should behave like Streck tubes if the workflow is properly executed.
* It is possible that some library prep inhibitors are present in blood. The most common exogenous/endogenous interfering substances (EDTA, Acetaminophen, Triglycerides, creatinine) have been tested on an the cfDNA workflow/chemistry and did not impact the library yields or conversion metrics.

**For the initial sonication step**, **can labs use DPSB** (**Dulbecco**'**s Phosphate**-**Buffered Saline**) **to dilute samples for shearing**, **or should the Resuspension Buffer** (**RSB**) **provided by Illumina be used**?

* Illumina recommends shearing in RSB.

**Can samples that are a mix of human and mouse DNA be run with this prep**?

* In theory, as long as the DNA is "clean" (eg, some silica column is used as part of the DNA purification) we would expect the methylation conversion to work. Analysis-wise, the customer will need to set up a custom genome of both human and mouse, will need to sort reads during alignment to the correct genome, and be aware of some amount of mapping ambiguity for any reads with high human/mouse sequence identity.
* The user may need to make a custom reference via the DRAGEN Reference Builder.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                              |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9945), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009945%20-%20Library%20Preparation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter, and the optional `goal` query parameter:

```
GET https://knowledge.illumina.com/library-preparation/epigenetics-library-prep/library-preparation-epigenetics-library-prep-faq-list/000009945.md?ask=<question>&goal=<endgoal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is optional and describes the broader end goal you are ultimately trying to accomplish on behalf of the user. GitBook uses it to tailor the answer towards what is most useful for that goal.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.