Workflow optimization by modulating SMB4 volume for Illumina DNA Prep with Exome 2.5 Enrichment v2

The standard volume for SMB4 (Streptavidin Magnetic Beads) usage with Illumina DNA Prep with Exome 2.5 Enrichment v2 is 200 µl, down from 250 µl used in the v1 kit. There is an optional step in the protocol to further reduce the SMB4 volume below 200 μl to a minimum of 100 μl. This may improve coverage of high-GC targets, but may also negatively impact SNV and indel recall rates. Users should determine whether to reduce volume based on sample type and variants of interest.

Example data for libraries prepared with 100 µl of SMB4 is available in the BaseSpace Demo Dataset: NovaSeq 6000 (S4): Illumina DNA Prep with Exome 2.5 Enrichment and Illumina DNA Prep with Exome 2.5 Enrichment v2 (HG001) with NovaSeq 6000 Sequencing System Software Suite v1.8.2., directly accessible here (note: BaseSpace login is required to view demo data).