> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/library-preparation/custom-third-party-library-prep/library-preparation-custom-third-party-library-prep-faq-list/000006127.md).

# Bead clean-up ratios used in the Illumina 16S Metagenomic Sequencing Library Preparation protocol

Below are the bead clean-up ratios used after each PCR step of the [16S Metagenomic Sequencing Library Preparation demonstrated protocol](https://support.illumina.com/documents/documentation/chemistry_documentation/16s/16s-metagenomic-library-prep-guide-15044223-b.pdf).

**PCR Clean-Up 1**

* This step uses purification beads to purify the 16S V3 and V4 amplicon away from free primers and primer dimer species.
* 25 µl of samples + 20 µl beads = **0.8X** beads-to-sample ratio

**PCR Clean-Up 2**

* This step uses purification beads to clean up the final library before quantification.
* 50 µl of libraries + 56 µl beads = **1.12X** beads-to-sample ratio

**Note**: As the beads-to-sample ratio changes, the length of fragments binding to the beads and left in solution also changes; the lower the ratio of beads-to-sample, the larger the final fragments will be at elution.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                              |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #6127), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000006127%20-%20Library%20Preparation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/library-preparation/custom-third-party-library-prep/library-preparation-custom-third-party-library-prep-faq-list/000006127.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.