> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/library-preparation/ampliseq-library-prep/library-preparation-ampliseq-library-prep-reference_material-list/000001238.md).

# Performing a spike-in of custom oligos into an AmpliSeq for Illumina Panel

Illumina offers custom spike-in panels that can be added into existing AmpliSeq for Illumina panels. AmpliSeq for Illumina spike-in panels are usually formulated at 50X concentration. See the steps below for adding a spike-in panel to an AmpliSeq for Illumina ready-to-use or custom panel.

* Most AmpliSeq panels are supplied at 2X concentration (working concentration), though not all. Consult the relevant reference guide to confirm panel concentration and volume usage information.
* Spike-in oligos are provided at 50X concentration.
* Combine the spike-in panel with the main panel 1:25.
  * Typically, this is adding 5 µl AmpliSeq spike-in panel to one tube of main panel pool; confirm in the relevant reference guide and volumes before combining oligos.
* Use combined spike-in + main panel pool following the standard library preparation protocol in the relevant reference guide.

Refer to the reference guide for the main AmpliSeq for Illumina panel being used for additional information on spike-in panel use.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                              |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1238), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001238%20-%20Library%20Preparation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/library-preparation/ampliseq-library-prep/library-preparation-ampliseq-library-prep-reference_material-list/000001238.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.