> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/library-preparation/ampliseq-library-prep/library-preparation-ampliseq-library-prep-faq-list/000001048.md).

# Frequently asked questions for AmpliSeq for Illumina On-Demand panels

AmpliSeq for Illumina On-Demand panels provide a choice of > 5000 pretested genes with known content relevant for inherited disease research, including hereditary cancer, primary immunodeficiency, hearing loss, and muscular dystrophy.

**Can the On-Demand panel be mixed with another custom panel?**

No, each custom panel (On-Demand, Gene, Hotspot) is its own, distinct panel.

**How many PCR cycles should be used to amplify the custom targets?**

Use the number of Primer Pairs as listed on the tube to inform the number of PCR cycles to use. The [reference guide](https://support.illumina.com/downloads/ampliseq-for-illumina-custom-and-community-panels-reference-guide-1000000036408.html) provides recommended PCR cycles based on primer pair/amplicon number.

**What are the minimum and maximum content (target) sizes?**

Amplicons: 24-15,000 amplicons

Genes: 1-500 genes

**What is the amplicon length?**

All On-Demand panels have a maximum amplicon size of 275 bp.

**Which genome builds are supported?**

Human hg19 only.

**What gene format is needed when adding targets in DesignStudio?**

HGNC format.

**How are the amplicons tested?**

Genes have been tested via sequencing on the NextSeq 550 with a 2x151 run configuration.

Not all gene combinations have been tested.

**What are the specifications for performance?**

As this is a custom AmpliSeq design, Illumina does not provide specifications for performance.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                              |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1048), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001048%20-%20Library%20Preparation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/library-preparation/ampliseq-library-prep/library-preparation-ampliseq-library-prep-faq-list/000001048.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.