Troubleshooting a dry system thermal pre-run check failure on the NovaSeq X Series

Background

For NovaSeq X Series instruments, a pre-run check is performed prior to sequencing. During this process, the thermal check in the dry check may occasionally result in a failure. In such cases, a message similar to the one shown below may be displayed on the control software.

TemperatureAirxxx with value ## DegC was greater than ## DegC.

Troubleshooting steps

  1. Check the laboratory environment.

    • Verify the room temperature. If it is higher than the acceptable range, request that it be lowered.

    • Check the filter condition:

      • Make sure the filter has been replaced appropriately.

      • Confirm that the filter is not blocked or clogged.

  2. Retry the Pre-Run Check, by selecting Retry.

  3. If the error occurs again:

      • Power cycle the instrument.

      • Reconfigure and start the run again.

  4. If the issue persists, contact Illumina Technical Support.