> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-troubleshooting-list/000010228.md).

# Troubleshooting a dry system thermal pre-run check failure on the NovaSeq X Series

**Background**

For NovaSeq X Series instruments, a pre-run check is performed prior to sequencing. During this process, the thermal check in the dry check may occasionally result in a failure. In such cases, a message similar to the one shown below may be displayed on the control software.

TemperatureAirxxx with value ## DegC was greater than ## DegC.

**Troubleshooting steps**

1. Check the laboratory environment.
   * Verify the room temperature. If it is higher than the acceptable range, request that it be lowered.
   * Check the filter condition:
     * Make sure the filter has been replaced appropriately.
     * Confirm that the filter is not blocked or clogged.
2. Retry the Pre-Run Check, by selecting Retry.
3. If the error occurs again:
   \*
   * Power cycle the instrument.
   * Reconfigure and start the run again.
4. If the issue persists, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                         |
| :---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #10228), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000010228%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-troubleshooting-list/000010228.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.