Troubleshooting a dry system thermal pre-run check failure on the NovaSeq X Series
Background
For NovaSeq X Series instruments, a pre-run check is performed prior to sequencing. During this process, the thermal check in the dry check may occasionally result in a failure. In such cases, a message similar to the one shown below may be displayed on the control software.
TemperatureAirxxx with value ## DegC was greater than ## DegC.
Troubleshooting steps
Check the laboratory environment.
Verify the room temperature. If it is higher than the acceptable range, request that it be lowered.
Check the filter condition:
Make sure the filter has been replaced appropriately.
Confirm that the filter is not blocked or clogged.
Retry the Pre-Run Check, by selecting Retry.
If the error occurs again:
Power cycle the instrument.
Reconfigure and start the run again.
If the issue persists, contact Illumina Technical Support.