> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-troubleshooting-list/000009387.md).

# Troubleshooting 'One or more unknown settings found' Errors during Run Planning on the NovaSeq X Series

**Background**

The NovaSeq X Series Control Software allows users to setup sequencing runs using the Run Planning user interface (UI). The Generate FASTQC Metrics is an optional analysis that will output various quality metrics for each sample.

If this option is selected for a DRAGEN App version that is not compatible with this option when configuring multiple DRAGEN Analyses, the following error will present:

![](/files/praGWExXBZdw5vg6BjIE)

One or more unknown settings found: GenerateFastqcMetrics

**Troubleshooting Actions**

This error message only presents for NovaSeq X Series instruments running Control Software v1.3 when setting up runs in Run Planning using multiple Analysis configurations where at least some of the Configurations use BCL Convert versions that are incompatible with the Generate FASTQC metrics option. See the following instructions to mitigate this issue:

1. From the Run Planning UI, cancel the run setup and return to the Runs page within the NovaSeq X Series Control Software.
2. Select **Create Run** to restart run setup using the Run Planning UI.
3. On the Run Settings page, DO NOT select either radio button for the **Generate FASTQC Metrics** and leave it at the default selection.
4. Proceed with the remaining selections in Run Planning to complete Configuration setup.
5. Repeat steps 1-4 for each DRAGEN Analysis Configuration.
6. After completing Run Setup and all Configurations, see the following steps edit the **Generate FASTQC Metrics** selection if needed:
   1. Under the Planned Runs tab, select the **Run** to edit.
   2. In the Run Review page, select **Edit** in the bottom-right corner.
   3. For the select the **Edit** link. next to the Configuration # that requires updating.
   4. Under Analysis Settings, select the appropriate radio button for the desired **Generate FASTQC Metrics** option.
   5. Select **Save**.
   6. Select **Save as Planned**.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9387), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009387%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter, and the optional `goal` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-troubleshooting-list/000009387.md?ask=<question>&goal=<endgoal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is optional and describes the broader end goal you are ultimately trying to accomplish on behalf of the user. GitBook uses it to tailor the answer towards what is most useful for that goal.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.