> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-troubleshooting-list/000007766.md).

# How to resolve 'An unexpected error has occurred with system integration' during initialization on the NovaSeq X Series Instruments

**Background**

Issues with the NovaSeq X Series Control Software connecting to the onboard Compute Engine (CE) result in a Critical System Error during initialization.

![](/files/EUDWGCQGty0d0tLYQmAn)

Critical system error: An unexpected error has occurred with system integration. Power cycle the instrument per system guide instructions and retry. If the issue persists, please contact Illumina Technical Support.

**Troubleshooting Actions**

1. Contrary to the error statement, **do not** power cycle the instrument.
2. Exit the NovaSeq X Series Control Software and wait 5 minutes, then restart the NovaSeq X Series Control Software.
3. If the issue persists, power cycle the instrument.
   1. For power cycling instructions, see Knowledge Article [How to Power Cycle the NovaSeq X Series Instruments](/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-reference_material-list/000007478.md) for details.
4. After logging into the Linux Operating System (OS), wait at least 10 minutes before starting the Control Software.
5. If the issue persists after power cycling, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #7766), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000007766%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter, and the optional `goal` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-troubleshooting-list/000007766.md?ask=<question>&goal=<endgoal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is optional and describes the broader end goal you are ultimately trying to accomplish on behalf of the user. GitBook uses it to tailor the answer towards what is most useful for that goal.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.
