> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-troubleshooting-list/000007765.md).

# How to requeue or perform a new analysis of existing data on the NovaSeq X Series

The NovaSeq X Series instruments can perform data analysis using the on-board DRAGEN server built into the platform. This guide discusses different ways to setup a new analysis of existing data, or requeue analyses using the integrated Illumina Run Manager (IRM) interface within the NovaSeq X Series Control Software.

Analysis of NovaSeq X Series runs in BaseSpace Sequence Hub requires a correctly formatted v2 sample sheet. Illumina strongly recommends using the [BaseSpace Run Planning](https://basespace.illumina.com/run-planning/run-settings) tool to generate a sample sheet for analysis to make sure the required fields are correct.

The requeue analysis option will **not** be available through the Control Software if:

* A sequencing run or maintenance wash is in progress on the instrument.
* The run data was deleted from the instrument.

**How to perform new local analysis of Manual or Cloud Runs**

Use the steps below to setup a new analysis from a Manual or Cloud run using run data stored on the NovaSeq X Series instrument.

1. From the Control Software home screen, select the **instrument icon** to open the global navigation menu, then select **Runs**.
2. In the Runs window, select the **Completed** tab.
3. Select the run of interest listed under the **Run Name** column to open the **Run Details** page.
4. Under the **Secondary analysis** header, select **Requeue Analysis**.
5. In the popup window, select **Requeue analysis from sample sheet**.
6. Select the **Upload** button, then navigate to the v2 sample sheet file.
7. Enter the file path to the intended Output Folder on the External Storage.
8. Enter a description for the requeue in the **Reanalysis** **Reason** field.
9. Select **Requeue Analysis** when finished.

**How to requeue analysis of a Local run**

Use the steps below to requeue analysis of data from a Local run on the NovaSeq X Series. This workflow is useful when editing sample index combinations or to add/remove DRAGEN Applications to the analysis workflow.

1. From the Control Software home screen, select the **instrument icon** to open the global navigation menu, then select **Runs**.
2. In the Runs window, select the **Completed** tab.
3. Select the run of interest listed under the **Run Name** column to open the **Run Details** page.
4. Under the Secondary analysis header, select **Requeue Analysis**.
5. In the popup window, select one of the following options.
   1. **Requeue analysis from this run** - Configure analysis using the Illumina Run Manager user interface.
   2. **Requeue analysis from sample sheet** - Configure analysis settings using a v2 sample sheet.
      1. Select the **Upload** button to select a v2 Sample sheet.
6. Enter the file path to the intended Output Folder on the External Storage.
7. Enter a description for the requeue in the **Reanalysis** **Reason** field.
8. Select **Requeue Analysis** when finished.
9. When selecting the **Requeue analysis from this run** option, perform any of the following actions as necessary:
   1. To include additional configurations, select **Add configuration**.
   2. To exclude a configuration from the requeue, select **Delete** next to the configuration name.
   3. To change configuration information, select **Edit** next to the configuration name.
10. Select **Requeue Analysis** when finished.

**How to requeue analysis performed on BaseSpace Sequence Hub**

* Analysis of a NovaSeq X Series run in BaseSpace Sequence Hub requires a correctly formatted v2 sample sheet. Using BaseSpace [Run Planning](https://basespace.illumina.com/run-planning/run-settings) is strongly recommended to generate a sample sheet for reanalysis to ensure the required fields are correct.
* To requeue a NovaSeq X Series run on BaseSpace, user can requeue through **Run Summary page in BaseSpace** or **use** [**BCL Convert App**](https://basespace.illumina.com/apps/15914899/BCL-Convert) **on BaseSpace**.

**Requeue a Cloud NovaSeq X Series run through Run Summary Page**

1. From the **BaseSpace Sequence Hub** home page, select **Runs** at the top of the page, then select the **sequencing run** from the list of completed sequencing runs.
2. From the **Summary** page, select **Status** option, then **Requeue**, then **Planned Run**.

![](/files/PrV12NDBAa2jrBWClbrh)

3. Select the radio button to select the **Sample Sheet** source to be opened in the **Run Planning** tool and select **Continue**.

![](/files/7a66asWsTSvb53kb56hB)

* If using a new Sample Sheet, the new sample sheet can be uploaded to BaseSpace.

4. If needed, edit the analysis configuration after importing Sample Sheet then select **Requeue** to start the new analysis.

**Requeue NovaSeq X Series run with the BCL Convert app**

1. From the **BaseSpace Sequence Hub** Home page, select **Projects** at the top of the page.

* To requeue an analysis within an existing Project, select the appropriate **Project** from the list.
* To create a new Project, select the **File** icon, then select **New** and follow the prompts to create a new **Project**.

2. Upload the corrected sample sheet to a project by selecting the **File** icon, then selecting **Upload**,then **Files**. Select the **Misc**/**Other file** type and import the v2 sample sheet CSV file.
3. Navigate to the [**BCL Convert app**](https://basespace.illumina.com/apps/15914899/BCL-Convert)and select **Launch Application** from the app page.
4. Select **Select Runs** to choose a run uploaded to the BaseSpace Sequence Hub account.
5. Under Override Sample Sheet, select **Select Dataset File(s)** to select the sample sheet uploaded in **Step 2**.

* Launching the **BCL Convert app** without a new sample sheet will use the previous sample sheet already in the run folder.

6. Select **Launch Application** to start the new analysis.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #7765), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000007765%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-troubleshooting-list/000007765.md?ask=<question>
```

The question should be specific, self-contained, and written in natural language.
The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.
