> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-reference_material-list/000009864.md).

# How to Power Down the NovaSeq X Series Instruments for a Planned Power Outage

**Background**

This article describes how to fully power down the NovaSeq X Series instruments in preparation for a planned power outage.

**Guidance**

This guidance is dependent on the Control Software version and accessibility.

* For Control Software v1.2.2 and above:
  1. From the NovaSeq X Control Software Home screen, select the **instrument** icon to open the global navigation menu, and then select **Settings**.
  2. From the Settings menu, select **Shutdown**.
  3. Wait until the power button on the right side of the instrument begins to pulse, then toggle the power switch on the BACK of the instrument to OFF.\
     ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-f3f29671e88fae6ab84aeaa9ca04416697b658e5%2Fimage1.png?alt=media)
  4. After power outage is resolved, toggle the **power switch** on the **back** of the instrument to ON, then wait for the power button on the right side to pulse.
  5. Once the power button begins to pulse, **press** the power button to start instrument boot cycle.
  6. Allow the instrument to boot, then login to the **Linux OS**.
  7. Once logged into the Linux OS, select **Activities** in the top-left corner, then select the **NovaSeq X Series Control Software** icon to launch the Control Software
* For Control Software v1.2 and below:
  1. From the NovaSeq X Control Software Home screen, select the **instrument** icon to open the global navigation menu, and then select **Settings**.
  2. From the Settings menu, select **Shutdown**.
  3. Wait at least 10 minutes for the instrument to completely shutdown. The monitor should be dark, with no text or cursors flashing in the top left corner.
  4. When the instrument computer has powered down and the monitor is dark, toggle the power switch on the back right side of the instrument to the OFF position.
  5. After power outtage is resolved, toggle the **power switch** on the **back** of the instrument to ON, then wait for the power button on the right side to pulse.
  6. Once the power button begins to pulse, **press** the power button to start instrument boot cycle.
  7. Allow the instrument to boot, then login to the **Linux Operating System (OS)**.
  8. Within the Linux OS, select **Activities** in the top left corner, then select the **NovaSeq X Series Control Software** icon and allow control software to initialize.
     * Note: The **Windows** key on the keyboard can be used to open the **Activities** menu and select the **NovaSeq X Series Control Software** icon.

If the instrument is expected to be shut down for 14 days or more, see **Extended Shutdown** guidance in the [Novaseq X Series support documentation](https://support-docs.illumina.com/IN/NovaSeqX/Content/IN/NovaSeqX/ShutDown.htm).

If unable to shutdown through the Control Software, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9864), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009864%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-reference_material-list/000009864.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.