> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/novaseq-6000/instrumentation-novaseq-6000-reference_material-list/000008435.md).

# Chemistry and Imaging on NovaSeq 6000

The NovaSeq 6000 uses two-channel chemistry and Patterned Flow Cells technology.

Two-channel SBS chemistry uses only two fluorescent dyes and images to identify all four bases, simplifying detection compared to four-channel SBS, which requires a separate dye for each base. (Figure1).

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-013a8e5447676ea5fc53b09713bfa0baccbfc94b%2Fimage1.png?alt=media\&token=a5dbd6d7-de9b-4f7b-91b7-8194efa057a3)

**Figure 1**: Two-Channel SBS Chemistry by schematic representation.

Images are taken using red and green filter bands. Thymines are labeled with a green fluorophore, cytosines are labeled with a red fluorophore, and adenines are labeled with both red and green fluorophores. Guanines are permanently dark (Figure 2). Nucleotides are identified by analysis of the different emission patterns for each base across the combination of Image 1 and Image 2 and are processed by image analysis software to identify which bases are incorporated at each well position.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-7e3587e44f5633a868b2e909356d6446fde8c412%2Fimage2.png?alt=media\&token=d1093d75-da91-41cb-a8b5-f963e01c3ba2)

**Figure 2**: Image analysis during SBS Chemistry with Two-Channel Detection.

The NovaSeq 6000 system uses a patterned flow cell with nanowells (Figure 3) and offers 4 different flow cell options, allowing output to scale to different applications and throughput levels.

Clustering and sequencing occur in the nanowells with a proprietary ExAmp chemistry to ensure that each nanowell in the flow cell generates a single clonal cluster.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-8c143335b1a227cc562cc0614d8463c01da63a66%2Fimage3.png?alt=media\&token=db79cc46-1287-482b-bfd1-f6eff77a8944)

**Figure 3**: Schematic representation of NovaSeq 6000 flow cell and the nanowells.

The NovaSeq 6000 Flow Cell specifications are further detailed in the table below.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-ff552602f899dd0de584762e0668d44ce5c80af2%2Fimage4.png?alt=media\&token=1683aee5-2d33-4f4f-bad6-9bc33463dea9)

For more information, see the Sequencing by Synthesis page [here](https://emea.illumina.com/science/technology/next-generation-sequencing/sequencing-technology/2-channel-sbs.html) and the NovaSeq 6000 Sequencing System Guide [here](https://emea.support.illumina.com/downloads/novaseq-6000-system-guide-1000000019358.html).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #8435), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000008435%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/novaseq-6000/instrumentation-novaseq-6000-reference_material-list/000008435.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.