How to check flow cell vacuum pressure on the NovaSeq 6000
Background
The NovaSeq Service Software (NVSS) can be used to check the vacuum pressure when the instrument or flow cell is the suspected cause of a vacuum issue.
To check the flow cell vacuum pressure:
Close the NovaSeq Control Software and launch NovaSeq Service Software (NVSS).
Once initialized, select the Flow Cell option on main screen.
Load wash flow cells on Sides A and B.
Pressure readings for the wash flow cell should be around -90 kPa and show a steady green check mark for the sensor.
If the pressure is low or the sensor is not steady, clean the flow cell and stage, and re-check pressure values.
If the vacuum pressure is within the expected range, perform a power cycle on the instrument and proceed with next run. See How to Power Cycle the NovaSeq 6000 System.
If vacuum pressure is not within expected range, contact Illumina Technical Support.
NovaSeq 6000 Control software v1.8 introduced an option to check vacuum pressure prior to run start. This can be done through the control software.
To check the flow cell vacuum pressure from NovaSeq control software v1.8 and above:
Open the NovaSeq Control Software (NVCS).
Select Tools from the dropdown menu on the upper left corner.

Select Flow Cell Vacuum.

Load wash flow cells on sides A and B.

Pressure readings for the wash flow cell should be around -90 kPa and show a steady green check mark for the sensor.
For any feedback or questions regarding this article (Illumina Knowledge Article #5390), contact Illumina Technical Support [email protected].
Last updated
Was this helpful?
