> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-troubleshooting-list/000001379.md).

# Troubleshooting a Specific Camera Disabled Failed to Detect Clusters error on the NextSeq 500/550

This notification indicates that the NextSeq 500/550 is unable to detect clusters at cycle 1 for a specific section, lane and/or surface(s) as opposed to all cameras/lanes/surfaces. The following article is for runs that have not stopped prematurely.

This notification is typically caused by cluster density issues or communication issues and, because the NextSeq 500/550 has multiple cameras, the run may progress to completion if other cameras/lanes/surfaces are performing.

**Potential error message:**\
![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-61e0051afc39b8b69b41636dfc68a5ffa68667ab%2Fimage1.jpg?alt=media\&token=ad08fa7a-ada7-4e00-94d1-3f88889589f7)

**Troubleshooting steps:**

1. Evaluate the run performance for other sections and lanes.
   * Compare the cluster density of other lanes to the recommended cluster density for the NextSeq 500/550 (170-220k/mm2 for well balanced libraries).
   * If the run is over/under clustered or shows poor performance, a sample or cluster density issue is suspected.
2. If run performance meets specifications\*, a communication issue is suspected; power cycle the instrument and perform the system checks.
   * For further information, search for Knowledge Base articles [How to Perform a NextSeq 500/550 Power Cycle](/instrumentation/general/instrumentation-general-reference_material-list/000001972.md) and [How to perform a NextSeq 500/550 system check](/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-troubleshooting-list/000001883.md).
   * If all system checks pass, proceed with using the instrument normally.
3. If issue persistently occurs with same camera and high or low cluster density are not suspected, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support) for further guidance.

For further information, search for Knowledge Base articles:

* [Cluster density guidelines for Illumina sequencing platforms using non-patterned flow cells](/instrumentation/general/instrumentation-general-reference_material-list/000001511.md)
* [Troubleshooting mid-run Camera Disabled error on the NextSeq 500/550](/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-troubleshooting-list/000001378.md)
* [Troubleshooting Cycle 1 Error, Camera Disabled, Failed to Detect Clusters on the NextSeq 500/550](/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-troubleshooting-list/000001681.md)

\*Note: See the support page for [NextSeq 500/550 Specifications](https://www.illumina.com/systems/sequencing-platforms/nextseq/specifications.html).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1379), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001379%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter, and the optional `goal` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-troubleshooting-list/000001379.md?ask=<question>&goal=<endgoal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is optional and describes the broader end goal you are ultimately trying to accomplish on behalf of the user. GitBook uses it to tailor the answer towards what is most useful for that goal.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.