> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/nextseq-1000-2000/instrumentation-nextseq-1000-2000-troubleshooting-list/000008307.md). Agents can also search and read this documentation through its [MCP server](https://knowledge.illumina.com/~gitbook/mcp).

# Investigating network data copy performance from Illumina Products on Linux OS

**Background**

Networks may have sub-optimal network and storage performance despite operating under the recommended network speed specifications such as 1 Gigabit per second (Gbps) or 10 Gbps connections. This can be due to many factors that are hard to diagnose such as competing traffic, hardware reliability, etc.

Looking at the output of the `ethtool` command can be used to verify that the instrument has negotiated the correct connection speed. However, it does not provide any practical information about potential network storage hardware bottlenecks.

**Troubleshooting actions**

The best way to measure actual networkperformance is through empirical testing of write performance to the network storage location. This can be achieved through the following steps:

1. Minimize the Control Software and then open the Linux Operating System (OS) Terminal app.
   * For NextSeq 1000/2000: select the **Applications** menu, then the **Terminal** app.
   * For NovaSeq X Series: select the **Activities** menu and then open **Terminal** app.
2. Input the following command, then press **Enter** to run it:
   * dd if=/dev/zero of=/path/to/networked/storage/test.txt bs=1M count=1024 oflag=direct

This command will write a test file titled "test" using data from a special Linux drive "/dev/zero". The file is written using 1 Mb blocks, for a total of 1024 blocks, amounting to a total write of 1 Gb. The total number of blocks can be increased or decreased to adjust the file size.

The command requires that the network storage is mounted and is accessible through the Terminal.

* For the NextSeq 1000/2000, see [Networking and output locations on the NextSeq 1000/2000](https://knowledge.illumina.com/instrumentation/nextseq-1000-2000/instrumentation-nextseq-1000-2000-reference_material-list/000003037.md).
* For the NovaSeq X Series, see [How to access network storage locations in Linux on the NovaSeq X Series](https://knowledge.illumina.com/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-reference_material-list/000008035.md).

The command returns the time taken to perform the write operation and the calculated data copy speed in Bytes/sec, which can be converted to bits/sec by multiplying this value by 8 bits/Byte.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-c5a6a1e9d2113227a27e0df9208239dd0260aa2a%2Fimage1.png?alt=media\&token=ac427244-f9c1-44aa-ac88-86c64d49e9f4)

**Figure 1**. Example using the `dd` command to test data copy speed to a network location.

Run the command several times to capture any variability in the network/storage performance.

When testing with the above method, it is important to note that it will not typically return the theoretical maximum speed. For example, a 10 GB switch may hit a maximum transfer speed of 6-7 Gb/s even under optimal conditions.

For recommended minimum network transfer speeds, see instrument-specific Site Preparation Guides.

More information, reference the applicable product documentation:

* [NextSeq 1000/2000 Product Documentation](https://support-docs.illumina.com/IN/NextSeq10002000/Content/IN/SitePrep/NextSeq1k_2k/NetworkConsiderations.htm)
* [NovaSeq X Series Product Documentation](https://support-docs.illumina.com/IN/NovaSeqX/Content/NetworkConnections.htm)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #8307), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000008307%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/nextseq-1000-2000/instrumentation-nextseq-1000-2000-troubleshooting-list/000008307.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.