> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-troubleshooting-list/000008538.md).

# Troubleshooting "Access to the path 'Global\Local Run Manager' is denied" on MiSeq

Users may encounter the error "Access to the path 'Global\Local Run Manager' is denied." while attempting to close or re-initialize the MiSeq Control Software.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-1e49b0881de7515e9f9005b6a7bd63353e1f76fa%2Fimage1.png?alt=media\&token=2cd2fda7-aea6-4c69-88cd-a29bf8974acc)

This error may be the result of failing to power cycle the instrument after updating the log-on credentials on the instrument.

**To resolve this error:**

1. Power Cycle the instrument. Instructions for this may be found in the Illumina Knowledge article [How to Power Cycle the MiSeq](https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-reference_material-list/000001894.md) or [How To Power Cycle the MiSeq Video](https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-reference_material-list/000005871.md).
2. Ensure all log-on credentials are configured as recommended in the Illumina Knowledge article [How to connect benchtop instruments and Local Run Manager to a network location](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000001934.md).

If the error persists, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #8538), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000008538%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-troubleshooting-list/000008538.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.