> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-troubleshooting-list/000007343.md).

# MiSeq RFID Bypass Code生成方法

\ <br>

MiSeq试剂都配有射频识别（RFID）标签以便对试剂信息进行准确记录。RFID的读取是在测序实验一开始由分别针对流动槽（Flow Cell），试剂卡盒（Reagent Cartridge）和PR2瓶3个不同位置的感应器记录追踪的。如果RFID没有被检测到，仍然可以通过手动输入bypass code来开始测序实验。Bypass code可以在登录MyIllumina账户后，利用MiSeq自助服务工具生成的。

&#x20;

MiSeq RFID Bypass Code生成方法：

1. 登录 [MyIllumina](https://dashboard.my.illumina.com/) .
2. 打开导航栏中的菜单**Tools**，然后选择**MiSeq Self-Service：**

&#x20; &#x20;

<img src="https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-89be768c34e9e78e0e57b577d1687d0b357c3e4e%2Fimage1.png?alt=media" alt="" height="314" width="358">

&#x20;&#x20;

3. 输入您的仪器的序列号（格式为：M####）
4. 在Type of Override Code下拉菜单列表中选择**RFID Override**
5. 选择**Get Code**生成code

<img src="https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-3f24e372a2b8264ee2fb5207dad1564aa11b0659%2Fimage2.png?alt=media" alt="" height="337" width="489">

由此获得的Bypass Code可以直接输入MiSeq Control Software中，7天内有效。同时MiSeq Control Software软件还会提示您输入试剂的条码（barcode）和产品编号（Part Number）信息。您可以在试剂标签上找到这些信息，具体位置如下图所示。其中条码用蓝色高亮标记，产品编号用黄色高亮标记。

![](https://supportassets.illumina.com/content/dam/illumina-support/images/bulletins/code1.png)

如果同一种试剂多次出现RFID读取失败的问题，可能是感应器出现了问题。对此持续性问题，请发邮件到大中华技术支持邮箱（[chinasupport@illumina.com](mailto:techsupport@illumina.com)）或者致电4000665835获得进一步帮助。

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #7343), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000007343%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-troubleshooting-list/000007343.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.
