> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-troubleshooting-list/000007341.md).

# 如何在MiSeq系统上重启RTA

&#x20;

RTA（实时分析，Real-Time Analysis）是一种集成软件，可在测序运行过程中分析流动池图像并计算相关指标，将这些图像转换为 BCL 文件。

仪器采集的预处理图像体积很大。因此，如果在未重启 RTA 的情况下磁盘空间不足以存储这些图像，测序运行可能会因此发生崩溃。

如果 RTA 停滞或网络路径中断，可能会出现以下错误提示。

![](https://knowledge.illumina.com/~gitbook/image?url=https%3A%2F%2F761066130-files.gitbook.io%2F%7E%2Ffiles%2Fv0%2Fb%2Fgitbook-x-prod.appspot.com%2Fo%2Fspaces%252FGM9W2DuBTgEXv1ClCm8H%252Fuploads%252Fgit-blob-5eb7ec4977bd2125a422eb2b514fcffc5b7a0c81%252Fimage1.png%3Falt%3Dmedia%26token%3Dbfb6aea3-ec4c-43ea-9f62-55cef3f45908\&width=300\&dpr=3\&quality=100\&sign=898d2546\&sv=2)

&#x20;

当出现该提示时，请按照以下说明操作：

1. 关闭该提示，或选择 **Close the program（关闭程序）**。
2. 在控制软件（Control Software）中，将该测序运行选择为 **Pause（暂停）**。
3. 手动停止并重新启动 RTA。在 MiSeq 上重新启动 RTA 的步骤如下：

   1. 确认 RTA 尚未完成：检查运行文件夹中是否不存在 **RTAComplete.txt** 文件\
      （路径：`D:\Illumina\MiSeq Temp\[run folder]`）。
   2. 点击 Windows **开始**按钮（屏幕左下角），搜索 **Task Manager（任务管理器）**。\
      *或者，按下键盘上的 **Ctrl + Alt + Delete** 打开任务管理器。*
   3. 在任务管理器中，选择 **Processes（进程）** 选项卡，右键单击 **RTA.exe**，然后选择 **End Process（结束进程）**。
   4. 要重新启动 RTA，双击 **RTAstart.bat** 文件，该文件位于\
      `D:\Illumina\MiSeq Temp\[run folder]\Config`。
   5. 在任务管理器的 **Processes（进程）** 选项卡中确认 **RTA.exe** 已存在。\
      *注意：一个小的黑色命令提示符窗口表示 RTA 正在运行，该窗口可以最小化。*

   &#x20;

&#x20;      4\. 在 RTA 重新启动后，在控制软件中选择 **Resume（继续）** 该测序运行。

如需了解有关 RTA 数据处理的更多信息，请搜索知识库文章 [MiSeq output and analysis folder quick reference guide](https://knowledge.illumina.com/instrumentation/miseq/instrumentation-miseq-reference_material-list/000005817)。

如果您在重启RTA中遇到任何问题，可以发邮件到大中华技术支持邮箱[chinasupport@illumina.com](mailto:techsupport@illumina.com)获得更多技术支持。

&#x20;

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #7341), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000007341%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-troubleshooting-list/000007341.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.
