> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-troubleshooting-list/000007340.md).

# 如何排查MiSeq平台上的第一个循环报错的原因？

\ <br>

MiSeq平台在第一个循环拍照时发生报错通常表明簇信号强度不够，仪器不能找到最佳的聚焦平面。这可能是由于簇密度过高、过低或簇生成失败造成的。

&#x20;

可能的报错信息包括：

* Best focus not found
* Best focus is too near the edge of range
* No usable signal found, it is possible clustering has failed
* Through-focus peak did not exceed SNR threshold
* Z Motor attempt to move outside soft limits
* New in MiSeq Control Software (MCS) v4.0: ThroughFocusScan CoarseFocus failed to find focus, and there is no fallback position.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-ff2997017dce798df6d23a2a411c1ca20446489b%2Fimage1.png?alt=media\&token=5dd2f5a6-d921-4b65-8f70-074c5e2a0994)

第一个循环报错可能由文库、运行设置、仪器或试剂问题引起。

&#x20;

可能的仪器原因包括：

* 参与簇生成或者第一个碱基化学反应试剂的液流问题
* 流动槽温控问题
* 光学系统问题（这种情况非常少）

可能的文库和试剂原因包括：

* 使用了过期或者储存条件不当的试剂
* 文库设计或者文库质量/定量问题
* 由于文库设计问题导致Read1引物杂交不好
* 使用了不合适的引物
* 簇密度过高或者过低
* NaOH的pH值
* 清洗槽的污染

第一个循环的错误表明仪器在第一个循环未能识别到足够信号，无法计算出flow cell上的适当焦点位置。MiSeq测序仪在首轮成像后出现的错误可能意味着簇强度不足，导致仪器无法找到最佳聚焦平面。

&#x20;

如果聚焦失败发生在测序过程中 (例如index的第一个循环或者Read1或Read4的中途) ，请搜索illumina knowledge Bases中的技术文档[Troubleshooting best focus errors occurring mid-run on the MiSeq](https://knowledge.illumina.com/instrumentation/miseq/instrumentation-miseq-troubleshooting-list/000001849), [Troubleshooting Z motor errors occurring mid-run ](https://knowledge.illumina.com/instrumentation/miseq/instrumentation-miseq-troubleshooting-list/000001850)和 [Troubleshooting No Intensity for Index Read on MiSeq](https://knowledge.illumina.com/instrumentation/miseq/instrumentation-miseq-troubleshooting-list/000001461)进行问题排查。

&#x20;

**故障排查步骤：**

首先可以按照[MiSeq System Guide](https://support.illumina.com/sequencing/sequencing_instruments/miseq/documentation.html)对仪器进行系统检查，确保仪器的流路、温度以及机械移动系统正常运行：

1. 按照仪器提示执行post run wash (运行后清洗)
2. 按照使用手册，重启机器
3. 去Manage Instrument（管理仪器）界面，点击System Check（系统检查）
4. 选择所有的运动测试，试剂管路冲洗，温度梯度以及液量测试。

当系统检查完成，点击屏幕左下方的Show Details（展开细节）按钮，得到检查结果的完整信息。如果是PR2以外的任何液量测试或者其它测试失败，请发邮件到[chinasupport@illumina.com](mailto:techsupport@illumina.com)去获得更多技术支持。

&#x20;

如果系统检查通过，则要考虑报错可能是由文库或者试剂问题导致的。我们需要分析以下常见问题：

* 检查试剂盒的有效日期以及正确储存条件。
* 确保文库设计适合在Illumina平台测序。
* 利用Illumina推荐的方法，检查文库质量和浓度。
* 确保自定义引物适合MiSeq 测序平台
* 确保自定义引物加入到试剂夹盒正确的孔位中（如果加入到已有引物的试剂孔中，sample sheet不需要修改，如果加入到自定义引物的试剂孔中，则sample sheet需要选择自定义引物）
* NaOH是新鲜稀释的且pH值大于12.5

如果文库或试剂没有明显问题，在文库中掺入20% PhiX重新测序。PhiX可以作为下一次测序簇生成的阳性对照，进而判断之前的错误是否是由文库问题导致的。

&#x20;

如果后续运行在第一个循环再次因错误停止，可以发邮件到Illumina技术支持邮箱[chinasupport@illumina.com](mailto:techsupport@illumina.com)获得更多技术支持。

&#x20;

更多资源链接:

[Denature and Dilute protocol generator](https://support-docs.illumina.com/IN/dnd-wizard/Content/dnd/landing.htm)

[MiSeq System Custom Primers Guide](https://support.illumina.com/downloads/miseq-system-custom-primers-guide-15041638.html)

&#x20;

**更多信息，请搜索 Knowledge Base中以下文档:**

[Spiking custom primers into the Illumina sequencing primers](https://knowledge.illumina.com/library-preparation/general/library-preparation-general-reference_material-list/000001542)

[How to achieve more consistent cluster density on Illumina sequencing platforms](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000001481)

[How to perform a System Check on the MiSeq](https://knowledge.illumina.com/instrumentation/miseq/instrumentation-miseq-reference_material-list/000001777)

[How To Power Cycle the MiSeq Video](https://knowledge.illumina.com/instrumentation/miseq/instrumentation-miseq-reference_material-list/000005871)

[Troubleshooting best focus errors occurring mid-run on the MiSeq](https://knowledge.illumina.com/instrumentation/miseq/instrumentation-miseq-troubleshooting-list/000001849)&#x20;

[Troubleshooting Z motor errors occurring mid-run on the MiSeq](https://knowledge.illumina.com/instrumentation/miseq/instrumentation-miseq-troubleshooting-list/000001850)&#x20;

[Troubleshooting No Intensity for Index Read on MiSeq](https://knowledge.illumina.com/instrumentation/miseq/instrumentation-miseq-troubleshooting-list/000001461) &#x20;

&#x20;

&#x20;

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #7340), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000007340%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-troubleshooting-list/000007340.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.
