> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-troubleshooting-list/000006801.md).

# Troubleshooting high cluster density on MiSeq

Overclustering leads to poor run performance, lower Q30 scores, the possible introduction of sequencing artifacts, and, counterintuitively, lower total data output. For a MiSeq v2 reagent chemistry, supported cluster density for well-balanced libraries is 1000-1200K/mm2. For MiSeq v3 reagent chemistry, the supported cluster density for well-balanced libraries is 1200-1400K/mm2. For low diversity libraries, Illumina recommends targeting a cluster density 30-40% beneath the optimal range for the chemistry version and platform used. The optimal amount of reduction required must be empirically determined.

Elevated cluster density does not result from instrument issues or sequencing consumable issues.

**Best practices for quantification:**

* Accurate quantification and proper quality check of next-generation sequencing libraries are key to a successful sequencing run.
* Different methods for quantification and quality control are recommended depending on the sequencing library kit being used.
* Consider the average size of the library. Because the clustering process preferentially amplifies shorter libraries in a mixture of fragments, large libraries tend to cluster less efficiently than smaller libraries.

**Best practices for sequencing low diversity libraries:**

* Nucleotide diversity is required for effective template generation (cluster mapping) on Illumina sequencing platforms and is important for generating high-quality data.
* Diversity is especially important during the first 4-7 cycles of the first sequencing read for MiSeq systems. The sequencing software uses images from these early cycles to identify the location of each cluster.
* A minimum of 5% PhiX is required for sequencing low base diversity libraries on MiSeq with v3 reagents.
  * The PhiX Control v3 Library has a diverse base composition (45% GC and 55% AT) that provides the balanced fluorescent signals that low diversity sample libraries lack during each sequencing cycle. This, in turn, assists with template registration and improves overall run quality.
* Reduce the library loading concentration to target a cluster density 30-40% beneath the optimal range for the chemistry version and platform used. The optimal amount of reduction required must be empirically determined.

**Additional Resources:**

[Cluster density guidelines for Illumina sequencing platforms using non-patterned flow cells](https://support.illumina.com/bulletins/2016/10/cluster-density-guidelines-for-illumina-sequencing-platforms-.html)\
[Cluster Optimization Overview Guide](https://support.illumina.com/downloads/cluster-optimization-overview-guide-1000000071511.html)

**For further information, search for Knowledge Base articles**:

* How to achieve more consistent cluster density on Illumina sequencing platform
* How much PhiX spike-in is recommended when sequencing low diversity libraries on Illumina platforms?
* Library quantification and quality control quick reference guide
* Best practices for library quantification
* Library quantification considerations when using qPCR
* Best practices for manually normalizing library concentrations

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #6801), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000006801%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-troubleshooting-list/000006801.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.