> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-troubleshooting-list/000005751.md).

# How to recover Local Run Manager v3 Account when locked out

When Account (User) Management is enabled, Local Run Manager (LRM) v3 allows creation of individual username(s) associated with the account (new users). When a new user is created, LRM will open a prompt to answer security questions. If a password is forgotten, users can recover access via selecting “Forgot your password” on the login screen and answering the security questions. Security questions can be reset anytime in the “My Account” page in LRM.

Users are required to either create or select 3 security questions from the default list. **Note**: the questions must be at least 20 characters and the answers must be at least 8 characters in length.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-d94e405b241481229b4d59fd176db4f5583da129%2Fimage1.png?alt=media\&token=0ee3cf6e-cb8f-419a-9d1a-90902e0a03c2)\
![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-91f4cc69b3a98857091e65289a2c51a27886ffeb%2Fimage2.png?alt=media\&token=8be1b840-347e-4d14-8012-118b9cdc5b3d)

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-54adf1b8f347d3c7477cbfe0a682eb4e4db6adfb%2Fimage3.png?alt=media\&token=81ec9ea8-ec42-4f50-b3f4-2585ae7408ca)\
![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-043b943632084d698cadafd93bb1feded85f3568%2Fimage4.png?alt=media\&token=eaca9510-0fc9-4463-850d-b68b3343480c)

Users will be locked out of their account if the incorrect password is entered repeatedly. The number of allowed attempts can be set by the Admin user - The default is 5. If a user exceeds the number of allowed attempts, the user account is locked. Only an Admin user can unlock the account or assign a temporary password.

Passwords will automatically expire after 90 days. LRM will prompt User/Admin to change password once the password has expired. The password expiration and user lockout parameters can be edited in the System Settings - Security tab.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-3fa3e9b5017c6f9370651ac6a286b243120812d1%2Fimage5.png?alt=media\&token=d6235181-0563-40a5-9713-022c7df98efc)

If locked out, admins and users can select “Forgot your password” to answer security questions and recover the account. Users will need to answer 2 questions correctly. Upon successfully answering the security questions, the user is prompted to create a new password.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-a47efb79e8889e25ec719f4cf0bdb443dd15caa3%2Fimage6.png?alt=media\&token=4b615fc0-0dd8-41ca-94f1-c4a5ec557863)\
![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-c39171d7ddeef9e1c12f3970ce39556fd410e75e%2Fimage7.png?alt=media\&token=5068d8fa-cbcf-4248-9eb0-dc924ba462d4)

If all Admin accounts get locked out, contact Illumina Technical Support.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #5751), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000005751%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-troubleshooting-list/000005751.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.