> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-troubleshooting-list/000002092.md).

# Troubleshooting mid-run shutdowns or reboots on the MiSeq

Users may encounter a MiSeq with the screen black or the Windows log in screen while a run was in progress. These mid-run instrument shutdowns and/or reboots may result from environmental or instrument-related root causes. These may be different from Blue Screen errors - for more information regarding a blue screen error on the MiSeq, review the article [Troubleshooting Windows OS Crashes or Blue Screen Errors on Illumina Instruments](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-troubleshooting-list/000001447).

To identify if the instrument spontaneously rebooted or shutdown, determine what was present on the instrument screen at the time the issue was discovered. The instrument screen will be at the home log-in screen if the system rebooted, and will display a black screen if the system shut down without rebooting.

**To troubleshoot an environmental power supply**:

1. Verify no interruption to power within the lab or power outage occurred. This may require contacting local facilities teams.
2. Verify instrument is connected to an Uninterruptable Power Supply (UPS). A UPS will provide several minutes of extra power, but will not sustain the instrument through longer periods of power outage.
   * For more information, review article [Uninterruptible Power Supply (UPS) Models by Instrument](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000005838)[.](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000005838)
3. Verify that no other equipment or instruments in lab were impacted.
4. Verify the power switch on the back of the instrument is toggled to ON.

If there were no unexpected interruptions to the power supply to the instrument and the power switch is toggled to the ON position, collect the following files and information from the instrument and contact Illumina Technical Support at <techsupport@illumina.com> for further assistance.

* Collect the Run cycle logs (located in D:\Illumina\MiSeq Temp\[Run\_Folder]\Logs). Include the InterOp folder, RunInfo.xml file, and RunParameters.xml file as well.
* Collect the MiSeq Control Software Logs (located in D:\Illumina\MiSeq Control Software).
* Collect the Windows Event Logs from the instrument as following instruction:
  1. Search for **Event Viewer** in the Start menu and open **Event Viewer**.
  2. Click on the down arrow to show the **Windows Logs**.
  3. Select **Application**, which will load a long list of errors, information, and warnings.
  4. Save this event log by right clicking on **Save All Events As…** on the far right and save as an event file (.evtx).
  5. Once the Application logs are saved, save the **Security** and **System** logs the same way as the Application logs (see steps 3 and 4 above).
* Provide a screenshot of the "Installed Products" Page in the Windows Control Panel.
  1\.
  1. To view Installed Products: Navigate to Control Panel > Programs and Features > Uninstall or Change a Program > Sort by "Installed On".
* Provide a screenshot of the" Windows Updates" page.
  1\.
  1. Windows 10: Start Menu > Settings > Update and Security > Windows Update > Update History > open to view if any recently installed updates.
  2. Windows 7: Start > Control Panel > Windows Update > "View update history" link.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #2092), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000002092%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-troubleshooting-list/000002092.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.