> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-troubleshooting-list/000001850.md).

# Troubleshooting Z motor errors occurring mid-run on the MiSeq

A Z-motor error occurring mid-run can be caused by instrument fluidics, debris, smudges on the flow cell, or library overclustering issues.

**ERROR:** "While trying to find focus, the Z stage has attempted to move outside its acceptable range."

**Troubleshooting steps:**

1. Confirm reagent kits were not expired at the time of the run.
2. Confirm the quality and quantification of the library were as expected and were checked in accordance with Illumina-recommended methods.
3. Contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/) with the following run files to review the library performance.
   1. From within D:\Illumina\MiSeqTemp\&lt;run folder>, send:\
      InterOp folder\
      RunParameters.xml\
      RunInfo.xml\
      Thumbnail\_Images\L001\C1.1 (cycle 1)\*\
      \*Send cycles 1-10 and the last 10 cycles before the run failed.
4. Power cycle\* the instrument. Reference[How to Power Cycle the MiSeq](/instrumentation/miseq-ruo/instrumentation-miseq-ruo-reference_material-list/000001894.md). This information is also available in video form. Refernce article [How To Power Cycle the MiSeq Video](/instrumentation/miseq-ruo/instrumentation-miseq-ruo-reference_material-list/000005871.md).
5. If the instrument initializes successfully, perform a System Check on the instrument. Reference [How to perform a System Check on the MiSeq](https://github.com/illumina-swi/illumina-knowledge/tree/master/docs/Instrumentation/MiniSeq/Reference_Material/000001777.md).
6. If the system check passes and Technical Support does not suspect a library-related issue, proceed with setting up the next run.
7. If Z-motor error recurs in the subsequent run, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1850), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001850%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter, and the optional `goal` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-troubleshooting-list/000001850.md?ask=<question>&goal=<endgoal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is optional and describes the broader end goal you are ultimately trying to accomplish on behalf of the user. GitBook uses it to tailor the answer towards what is most useful for that goal.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.