For the complete documentation index, see llms.txt. This page is also available as Markdown.

Troubleshooting Cavro Pump Error on MiSeq

A MiSeq Cavro Pump error can result from an instrument hardware issue or software communication issue on the MiSeq. Potential hardware-related causes include pump, pump controller, or syringe failure.

Error: Cavro pump error: Volume will exceed syringe capcity.Error in LanePump. Invalid argument. Command = '1V400IP1500R\r', reply = 'c'

Troubleshooting Steps:

  1. Power cycle the MiSeq.

    1. Note: If a run is in progress, it will NOT be able to be resumed.

  2. Perform a post-run wash to ensure fluidics are operating normally.

  3. If post-run wash or volume test is successful, proceed with next run.

  4. If the instrument fails to initialize or there is an error during the post-run wash, contact Illumina Technical Support for further assistance.

Note: If a Lane pump error does not include the Cavro pump error, refer to article Troubleshooting Lane Pump Errors on the MiSeq.

For any feedback or questions regarding this article (Illumina Knowledge Article #1454), contact Illumina Technical Support techsupport@illumina.com.

Last updated

Was this helpful?