Troubleshooting Y Motor Errors on the MiSeq
The following MiSeq Y motor errors usually stop the run and indicate communication problems with the stage or problems with stage hardware:
Y Motor: Closed-loop move cannot reach target position: Start = 75.4150390625 mm, Target = 75.40472 mm, Distance = -0.0103190625000025 mm, Stopped at 74.42236328125 mm, ExceptionMessage =
The FPGA reported an error while executing command "YMVTO 154429"; error message: "Error: too many steps needed"
Error: Timeout of 60.000 seconds exceeded while waiting for FPGA command "YMVTILE 1327" to complete


Troubleshooting Steps:
Perform the post-run wash, if the run stops or after the run is completed.
Power cycle the MiSeq. Reference article of How to Power Cycle the MiSeq or How To Power Cycle the MiSeq Video
If no issues with re-initialization, perform the Y Stage test 5 times. Reference article of How to perform a System Check on the MiSeq
Make sure that a used flow cell is loaded on the instrument (the same wash flow cell should be okay).
Select Y Stage test.
MiSeq Control Software (MCS) v 4.0: From Main Menu, select System Check.
MCS 3.1 and earlier: From Welcome Screen, select Manage Instrument, select System Check.
Follow the prompts on the screen to start the test, and repeat the test 4 more times (5 times total).
If all Y Stage tests pass continue with normal sequencing.
If the system fails to initialize or any of the 5 replicates of the Y Stage test fail, contact Illumina Technical Support for further assistance.
For any feedback or questions regarding this article (Illumina Knowledge Article #1452), contact Illumina Technical Support techsupport@illumina.com.