> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-troubleshooting-list/000001452.md).

# Troubleshooting Y Motor Errors on the MiSeq

The following MiSeq Y motor errors usually stop the run and indicate communication problems with the stage or problems with stage hardware:

* Y Motor: Closed-loop move cannot reach target position: Start = 75.4150390625 mm, Target = 75.40472 mm, Distance = -0.0103190625000025 mm, Stopped at 74.42236328125 mm, ExceptionMessage =
* The FPGA reported an error while executing command "YMVTO 154429"; error message: "Error: too many steps needed"
* Error: Timeout of 60.000 seconds exceeded while waiting for FPGA command "YMVTILE 1327" to complete

  ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-38d77c353d67a796f544fc5931614b614af9057b%2Fimage1.png?alt=media\&token=787f2182-865c-4cd6-89b3-2bc2a9f1f217)

  ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-b03accaef0186875d84acfa93352916168fc0c89%2Fimage2.png?alt=media\&token=14e0edcf-1b78-4f60-a92e-2cf729b22a64)

**Troubleshooting Steps**:

* Perform the post-run wash, if the run stops or after the run is completed.
* Power cycle the MiSeq. Reference article of [How to Power Cycle the MiSeq](https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-reference_material-list/000001894.md) or [How To Power Cycle the MiSeq Video](https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-reference_material-list/000005871.md)
* If no issues with re-initialization, perform the **Y Stage test 5 times.** Reference article of [How to perform a System Check on the MiSeq](https://github.com/illumina-swi/illumina-knowledge/tree/master/docs/Instrumentation/MiniSeq/Reference_Material/000001777.md)
  * Make sure that a used flow cell is loaded on the instrument (the same wash flow cell should be okay).
  * Select Y Stage test.
    * MiSeq Control Software (MCS) v 4.0: From Main Menu, select System Check.
    * MCS 3.1 and earlier: From Welcome Screen, select Manage Instrument, select System Check.
  * Follow the prompts on the screen to start the test, and repeat the test 4 more times (5 times total).
* If all Y Stage tests pass continue with normal sequencing.
* If the system fails to initialize or any of the 5 replicates of the Y Stage test fail, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/) for further assistance.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1452), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001452%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-troubleshooting-list/000001452.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.
