> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-troubleshooting-list/000001444.md).

# Troubleshooting FPGA Timeout Errors on the MiSeq

Field-Programmable Gate Array (FPGA) is the circuit board that translates the commands from the instrument control software into action by an instrument component.

In simpler terms, the FPGA mediates communications between the subsystems of the MiSeq.

If the FPGA loses communication with a subsystem, the MiSeq Control Software (MCS) will show a software message mentioning the FPGA timeout.

* For example, if there is a command from the control software to aspirate a certain volume of reagent, the FPGA board translates that command into actual aspiration.
* Once the aspiration occurs, the board receives feedback that the component was responsive, which enables the next command to proceed.

These communication issues can be transient or can point to problems with instrument hardware.

**Example of Error Messages:**

* The FPGA reported an error while executing the command "YHOME"; error message: "ERROR: Home sensor never tripped error."
* WRN: FPGA: FPGA: Expected no response from FPGA: command = EXLED1ON; Response =
* ERR: Timeout of 60.000 seconds exceeded while waiting for FPGA command
* WRN: Y Motor: Y Motor: UpdatePositionAndLimits failed: FPGA: Unable to write 'YRDPOS\n': Exception = 'The write timed out.'
* WRN: Y Motor: Y Motor: MoveAbsScaled failed: Y Motor: UpdatePositionAndLimits failed: FPGA: Unable to write 'YRDPOS\n': Exception = 'The write timed out.'

  ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-ef97af7494dfa1725ed88c492768182cf372ee8e%2Fimage1.png?alt=media\&token=2269a75c-c09b-4c92-83d7-76e69e1e40c3)

**Troubleshooting steps**:

1. E-mail Illumina Technical Support at <techsupport@illumina.com> with the information from the section below titled **Information to Provide to Tech Support.**
2. Complete the post-run wash.
3. Power cycle the MiSeq (note that if a run is in progress, it will not be able to be resumed).
   1. Close MiSeq Control Software shut down Windows, and turn off the instrument using the power switch on the back panel.
   2. Leave the instrument off for 5 minutes to allow all parts to completely power down.
4. Turn on the instrument using the power switch on the back panel. The control software will initialize on its own once Windows fully loads.
   1. For further information, see [How to Power Cycle the MiSeq](https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-reference_material-list/000001894.md) or [How To Power Cycle the MiSeq Video](https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-reference_material-list/000005871.md).
5. If the control software fails to initialize, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support).
6. If the instrument initializes without error, proceed with the next run.

**Information to provide to Tech Support:**

1. System uptime: This can be seen from Task Manager - Performance tab.
2. Software uptime: Available on the 'MiSeq Control Software' About screen (the time restarts when software is launched).
3. MiSeq Control Software (MCS) version: Available on the 'MiSeq Control Software' About screen.
4. FPGA version: Available on the 'MiSeq Control Software' About screen.
5. MiSeq Control Software logs: D:\Illumina\MiSeq Control Software\Logs
6. Run cycle logs, if applicable: D:\Illumina\MiSeqAnalysis\&lt;run folder>\Logs folder
   * Also available in D:\Illumina\MiSeqTemp\&lt;run folder>\Logs folder
7. Instrument usage rate: On average, how many runs were performed in the last month?
8. Run type, if applicable, when the error happened: Chemistry version and length of the run.
   * example: v2 500 cycle kit, 251 x 8 x 8 x 251

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1444), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001444%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-troubleshooting-list/000001444.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.