> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-troubleshooting-list/000001428.md).

# Troubleshooting an Object Reference Error on MiSeq

**Background**

An "Object reference not set to an instance of an object" notification is a generic Windows framework error that may occur during any stage of sequencing or when the system is idle. The notification does not necessarily stop a sequencing run and it is often caused by communication issues between instrument subsystems or network connectivity problems.

**Note:** In some cases, this notification may co-occur with SIPDOWN errors. If the error occurs right at run start and references SIPDOWN, see article [Troubleshooting SIPDOWN Errors on the MiSeq](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-troubleshooting-list/000001456.md).

**Troubleshooting steps:**

1. If a run is in progress, let it continue and monitor run performance.
2. Power cycle the MiSeq upon run completion, following the steps in [How to Power Cycle the MiSeq](https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-reference_material-list/000001894.md).
   1. **Note:** If a run is in progress, it cannot be resumed after the power cycle.
3. If the system re-initializes successfully, proceed with using the MiSeq system for sequencing.
4. If the instrument fails to initialize or any of the sequencing metrics look suboptimal, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support).
   1. For guidance on evaluating the sequencing run performance, see the article [Does my sequencing run look good?](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000001922.md)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1428), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001428%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-troubleshooting-list/000001428.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.