> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-reference_material-list/000002256.md).

# How to configure MiSeq Reporter to use a network location as the default output folder

The MiSeq Reporter software is responsible for generating FASTQ and other analysis files from the sequencing run data. See the following steps to configure the MiSeq Reporter software to output these files directly to a network storage location.

1. Make sure the MiSeq instrument control computer is on the same domain as the server that will be used as the network storage location.
   1. To check this, go to the Windows Start menu and right-click Computer.
   2. Select Properties and check the Domain name under 'Computer name, domain, and workgroup settings'.
2. Change the MiSeq Reporter Service to Run as `sbsuser`:
   1. On the MiSeq, from the Windows Start menu, enter `run` in the Search field and press the Enter key.
   2. In the Run dialog box, enter `services.msc`
   3. In the Services window, scroll down to locate MiSeq Reporter, right-click, and select Properties.
   4. Select the 'Log On' tab.
   5. Click 'This account' and enter `.\sbsuser`
   6. Enter the sbsuser password in the 'Password' field.
   7. Select 'Apply' then 'OK'.
   8. Back in the Services app, right-click on MiSeq Reporter and select 'Restart'.
3. Consult with the local IT team to make sure that the local `sbsuser` account on the MiSeq has read/write privileges to the desired output folder on the server.
   1. To test read access on the MiSeq, open Windows Explorer and enter the fully qualified Universal Naming Convention (UNC) path to the folder in the address bar, then press Enter.
   2. To test write access on the MiSeq, open Windows Explorer and navigate to the network location, then create a new folder or text file.
   3. If there are errors with either of these actions, consult with the local IT team to resolve the read/write permission issues.
4. In the MiSeq Control Software (MCS), set the MiSeq Output folder to the location of the Output folder on the Windows server.
   1. On the MiSeq Welcome screen, select Run Options.
   2. On the Run Options screen, select the Folder Settings tab.
   3. Enter the location of the network folder using the fully qualified UNC path.
      1. Example: \server\_ipaddress\share\output\\
   4. Click Save and Return.

**Note**: If MiSeq Reporter software is updated or reinstalled, repeat all the steps listed in these instructions.

If additional assistance is required, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #2256), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000002256%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-reference_material-list/000002256.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.