> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-reference_material-list/000002031.md).

# How to stop and restart RTA on the MiSeq

RTA (Real-Time Analysis) is an integrated software that analyzes flow cell images and calculates metrics during a sequencing run, converting those images into BCL files.

The pre-processed images captured by the instrument are large in size. As such, if RTA is stalled or stopped and is not restarted the run may terminate.

If RTA stalls or the network path is down, the following error message may appear:\
![](/files/ot3jRfEO9U1dLJz3nVcO)

When this notification occurs, follow the instructions below to restart RTA:

1. Close the notification or select **Close the program**.
2. Select **Pause** on the run in the Control Software.
3. Manually stop and restart RTA.

* **To restart RTA on the MiSeq:**\
  1\. Confirm that RTA has not completed by verifying that there is no RTAComplete.txt file present in the run folder (D:\Illumina\MiSeq Temp\[run folder]).\
  2\. Select the Windows **Start button** (bottom left of the computer screen), and search for **Task Manager**.\
  \* Alternatively, press Ctrl+Alt+Delete on the keyboard to open Task Manager.\
  3\. In Task Manager, select the **Processes** tab, right-click on **RTA.exe**, and select **End Process**.\
  4\. To restart RTA, double-click the **RTAstart.bat** file, located in D:\Illumina\MiSeq Temp\[run folder]\Config.\
  5\. Confirm that RTA.exe is present in the Task Manager Processes tab.\
  \* Note: A small black command prompt window indicates that RTA is running, and it can be minimized.

4. Once RTA has been restarted, select **Resume** on the run in the Control software.
5. If error repeats, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support) for assistance.

For further information on data processing by RTA, search for Knowledge Base article [MiSeq output and analysis folder quick reference guide.](https://knowledge.illumina.com/instrumentation/miseq/instrumentation-miseq-reference_material-list/000005817)

For additional guidance, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #2031), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000002031%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter, and the optional `goal` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-reference_material-list/000002031.md?ask=<question>&goal=<endgoal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is optional and describes the broader end goal you are ultimately trying to accomplish on behalf of the user. GitBook uses it to tailor the answer towards what is most useful for that goal.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.