> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-reference_material-list/000001678.md).

# File paths for the MiSeq

**MiSeq\*\*\*\*File Paths**

There are three locations for the run folders from each run. Each location has overlapping content, though they are not mirrors of each other.

1. **MiSeq Temp Folder**: Temporary working folder for MiSeq Control Software (MCS) and Real Time Analysis (RTA) software. Contents of this folder are deleted automatically after seven days.
2. **MiSeq Analysis Folder**: As primary analysis files are generated, RTA copies files back to the Temp folder and populates the Analysis folder. When the primary analysis is complete, MiSeq Reporter (MSR) or Local Run Manager (LRM) accesses the Analysis folder to begin secondary analysis. All files written to the Analysis folder are copied to the Output folder by RTA (Windows 7 - MCS v3.1 and older) or Universal Copy Service (UCS) (Windows 10 - MCS v4.0).
3. **MiSeq Output Folder (or network path)**: RTA or UCS copies files from the Temp folder to the Output folder.

**The default run folder name uses the following format**:

YYMMDD\_\[InstrumentSerialNumber]\_\[Run Number]\_A\[FlowCellBarcode]

**Run Folder Architecture\***

D:\Illumina\MiSeq Temp(or MiSeq Output or MiSeq Analysis)\[Run\_Folder]

* Runinfo.xml
* Runparameters.xml
* InterOp folder
* SampleSheet.csv
* Cycle Logs:
  * D:\Illumina\MiSeq Temp\[Run\_Folder]\Logs
* Thumbnail Images:
  * D:\Illumina\MiSeq Temp\[Run\_Folder]\Thumbnail\_Images\L001\C# (folder(s) listed by cycle) (Not present in Analysis folder, but present in Temp and Output folders)
* Offsets folder:
  * D:\Illumina\MiSeq Temp\[Run\_Folder]\Data\Intensities\Offsets
* RTA Logs:
  * D:\Illumina\MiSeq Temp\[Run\_Folder]\Data
* Recipes:
  * D:\Illumina\MiSeq Temp\[Run\_Folder]\Recipe

**Analysis Files**

* bcl, stats and locs files:
  * D:\Illumina\MiSeq Analysis\[Run\_Folder]\Data\Intensities\BaseCalls
* cif files:
  * D:\Illumina\MiSeq Analysis\[Run\_Folder]\Data\Intensities\L001\C
* cif files (alternative location):
  * D:\Illumina\MiSeqTemp\Processed\L001\Cx.1
* DemultiplexSummaryF1L1.txt for MiSeq Reporter:
  * D:\Illumina\MiSeqAnalysis\[Run\_Folder]\Data\Intensities\BaseCalls\Alignment\\
* DemultiplexSummaryF1L1.txt for Local Run Manager (LRM) v2, v3 and v4:
  * D:\Illumina\MiSeqAnalysis\[run folder]\Alignment\_#\[date and time folder]
* FASTQ files for MiSeq Control Software (MCS) v2.5 and v2.6:
  * D:\Illumina\MiSeqAnalysis (or MiSeqOutput)\[RunFolderName]\Data\Intensities\BaseCalls
* FASTQ files for MCS v3.1/LRM v2, MCS v4.0/LRM v3 and MCS v4.1/LRM v4:
  * D:\Illumina\MiSeqAnalysis (or MiSeqOutput)\[RunFolderName]\Alignment\_1\[autogenerated folder with date and time when analysis started]\Fastq
* MCS v3.1 and older LRM Analysis Logs:
  * D:\Illumina\MiSeqAnalysis\[RunFolderName]\Alignment\_#\[date\_time]

**Maintenance Logs**

D:\Illumina Maintenance Logs (available files inside the folder):

* Green LED Log.csv
* Red LED Log.csv
* Temperature Log Chiller.csv
* Temperature Log Flow cell #1.csv
* WashLog.xml

**Other important files**

* MCS Logs
  * D:\Illumina\MiSeqControlSoftware
* MiSeqSoftware.Options.cfg file:
  * Windows 7: D:\Illumina\MiSeqControlSoftware
  * Windows 10: C:\ProgramData\Illumina\MiSeqControlSoftware
* System check results:
  * D:\DiagResults\[Folder titled with the date the check is performed]
* Wash log:
  * D:\Illumina Maintenance Logs
* WashStatus.xml file:
  * MCS v3.1 and v2.6:
    * D:\root directory
  * MCS v4.0/v4.1:
    * C:\ProgramData\Illumina\MiSeq Control Software
* BaseSpace Broker folder:
  * D:\Illumina\basespacebroker
* MiSeqOverride.cfg
  * Windows 7: C:\MiSeq Config
  * Windows 10: C:\ProgramData\Illumina\MiSeq
* Runcompletionstatus.xml
  * D:\Illumina\MiSeqOutput\[Run\_Folder]

**MiSeq Control Software v4 specific file paths**

* UCS Logs and Run folders
  * C:\ProgramData\*\Illumina\UniversalCopyService
* LRM Logs folder
  * C:\ProgramData\*\Illumina\LocalRunManager

**\*Note:** The ProgramData folder may be hidden; to un-hide files, click on the View tab in the file explorer window and ensure Hidden items is checked.

**Additional Resources**:\
\*For an in-depth description of each folder/function, review the [MiSeq Run Folder Quick Reference Card](https://support.illumina.com/downloads/miseq_run_folder_quick_reference_card_15034791.html) (MCS v2.6 and older).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1678), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001678%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-reference_material-list/000001678.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.