> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-reference_material-list/000001676.md).

# Recommendations for optimal cluster density and PhiX loading concentrations on the MiSeq

Cluster density is an important factor in optimizing sequencing data quality and yield. The list below includes the recommended raw cluster densities for balanced libraries (such as the PhiX control library, or other balanced/base diversity libraries, like whole genome libraries).

**MiSeq Reagent Kit v2**: **Standard**, **Micro**, **and Nano kits**

* Optimal Cluster density: 1000 - 1200 K/mm2 (for a well-balanced library).
* PhiX loading concentration: 12.5 pM.

**MiSeq Reagent Kit v3**: **Standard kit**

* Optimal Cluster density: 1200 - 1400 K/mm2 (for a well-balanced library).
* PhiX loading concentration: 20.0 pM.

**Best practices for sequencing low diversity libraries**:

* When running low base diversity or unbalanced libraries, Illumina recommends reducing the library loading concentration to **target a cluster density 30**-**40% below the ranges above** for the chemistry version and platform used. The optimal amount of reduction required must be empirically determined.
* To compensate for low base diversity in libraries, Illumina recommends **spiking in a minimum of 5% PhiX Control v3 Library for sequencing**. PhiX has a diverse base composition (45% GC and 55% AT) that provides the balanced fluorescent signals that low diversity libraries lack during each sequencing cycle.

For further information regarding PhiX on other Illumina sequencers, search for Knowledge Base article [How much PhiX spike-in is recommended when sequencing low diversity libraries on Illumina platforms?](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000001527.md)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1676), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001676%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-reference_material-list/000001676.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.