> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-reference_material-list/000001244.md).

# How to set up the sample sheet for a PhiX validation run on the MiSeq using IEM

**Background**

* A PhiX validation run confirms proper hardware and software performance of the instrument.
* The Illumina PhiX control library is a well-balanced genome with relatively equal representation of A, T, G, and C nucleotides.
* PhiX lacks an index and is not an appropriate tool for assessing Index Read performance.
* Below are instructions for setting up the sample sheet for a PhiX validation run for the MiSeq system using [Illumina Experiment Manager (IEM) v1.19.1](https://support.illumina.com/sequencing/sequencing_software/experiment_manager/downloads.html?langsel=/us/).

**Procedure**

\*\*1.\*\*Open Illumina Experiment Manager and select **Create Sample Sheet**

**2**.Select the **MiSeq** instrument and select **Next**

\*\*3.\*\*Select **Other**

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-5051fef6875622be290afe5fdbd2c1b84652ba99%2Fimage1.png?alt=media\&token=21abea74-ffc2-49d5-80ab-bb031b80678d)

\*\*4.\*\*Select **FASTQ Only** and then **Next**

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-3855d71bd4a03074d66a9c7152fcaacd59cd3250%2Fimage2.png?alt=media\&token=cab7c034-c341-45a8-885b-f3b66443079e)

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-87255658f19ef0e9ef88d9b02543e4519ebb70a8%2Fimage3.png?alt=media\&token=180e1dbd-e32b-49c9-9791-076e04b5a872)\
![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-83759e6d5b67b3905dd589c89d933f13db662a79%2Fimage4.png?alt=media\&token=7a83d955-05ac-4370-8320-bd18622e3c92)

Run length and parameters vary depending on troubleshooting needs and reagent chemistry version. The recommended minimal read length is 26 cycles for Read 1 and Read 2 if using a MiSeq v2 50 cycle kit.

**Note**

* Some older versions of Illumina Experiment Manager do not have separate Library Prep Workflow and Index Adapters dropdown menus.
  * For versions with only one Library Prep Kit dropdown menu, select “TruSeq LT” as the workflow.

For more information about Illumina Experiment Manager and preparing PhiX for the MiSeq sequencing platform, see the following resources:

* [How to set up a PhiX validation run on MiSeq using software v2.6 and earlier](https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-reference_material-list/000001884.md)
* [Illumina Experiment Manager User Guide](https://support.illumina.com/downloads/illumina-experiment-manager-user-guide-15031335.html)
* [MiSeq System Denature and Dilute Libraries Guide](https://support-docs.illumina.com/IN/dnd-wizard/Content/dnd/landing.htm?ins=miseq)
* [PhiX loading concentrations for validation runs on Illumina sequencing platforms](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000001536.md)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1244), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001244%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-reference_material-list/000001244.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.
