> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-faq-list/000003094.md).

# Does MiSeq verify the number of cycles in a planned run against the cycles supported by the reagent kit?

**For MiSeq Control Software v3.1 and below:**

The MiSeq Control Software does not cross-reference the total number of cycles programmed in the run set up compared to the number of cycles a reagent kit can support. If a run is set up with a greater number of cycles than the reagent kit supports, the MiSeq Control Software allows the run to proceed, although run performance may be impacted. Runs may stop with Best Focus or Z-motor soft focus limit errors.

**For MiSeq Control Software v4.0 and above:**

With the introduction of MiSeq Control Software v4.0, the Software checks the total number of cycles to ensure it is less than or equal to the total cycles supported by the loaded reagent kit. This takes into account the 7 dark cycles required for Index Read 2 if the workflow uses dual index sequencing.

If the planned cycle count exceeds the maximum number of cycles supported by the reagent kit, the control software displays the following warning message after the pre-run checks:

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-8377ff97d6df70d4284bf211173aa28283911a9a%2Fimage1.png?alt=media\&token=887aacae-295e-459b-ae86-fa2ed2a3bf6e)

`"Correct the Following Issues: Cycles exceed reagent kit limit"`

If this error presents, review the cycle numbers for each read, and verify that the loaded kit supports the number of cycles required for the desired workflow.

For more information, refer to the following Illumina Knowledge Articles below:

[How many cycles of SBS chemistry are in my kit?](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000007002.md)\
[Maximum read length for Illumina sequencing platforms](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000002826.md)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #3094), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000003094%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-faq-list/000003094.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.