Which controls should be used when sequencing low-diversity libraries on a MiSeq?

Illumina recommends adding PhiX at 1% to all sequencing runs as an internal control.

For low-diversity libraries, spike in higher percentages of phiX to add base diversity to the run. The percentage of PhiX to depends on the diversity of the library, and requires optimization.

  • Monotemplates may require 30% PhiX.

  • Real Time Analysis (RTA) v1.17.28 or later requires a minimum of 5% PhiX spike-in for low-diversity libraries, when using MiSeq v3 reagents for sequencing.

  • MiSeq v2 runs require higher phiX spike in than MiSeq v3 runs.

For any feedback or questions regarding this article (Illumina Knowledge Article #2522), contact Illumina Technical Support techsupport@illumina.com.