# Troubleshooting Bad Request Errors When Adding Output Folder on MiSeq i100 Series

**Background**

The MiSeq i100 Series will test both read and write permissions when checking connectivity to a desired Output Folder on a configured External Storage. If the check is unable to complete the write test for the selected Output Folder, the following error message will present:

![](/files/LIYHoRmhxVcvgjwNMEDN)

Bad Request. Connectivity check failed to write to selected customer storage bookmark location. Check permissions issues with the selected output folder.

**Troubleshooting Actions**

This issue is caused by incorrect write permissions configured for the output folder on the configured network External Storage location. Consult the local IT team responsible for permissions management and work with them to ensure that the instrument has the appropriate **read and write** permissions for the intended Output folder.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                         |
| :---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #10126), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000010126%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions: Querying This Documentation

If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-troubleshooting-list/000010126.md?ask=<question>
```

The question should be specific, self-contained, and written in natural language.
The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.