Troubleshooting “The loaded dry cartridge is not supported for sequencing” error on the MiSeq i100 S
Background
MiSeq i100 Series instruments require Control Software v1.1 to MiSeq i100 25M Reagent Kit (1000 cycles), as well as all MiSeq i100 Plus 50M and 100M Reagent Kits. The error “The loaded dry cartridge is not supported for sequencing. Load a compatible consumable for sequencing.” displays to users attempting to load one of these kits on a MiSeq i100 Series instrument with v1.0 control software installed.

Troubleshooting Actions
Navigate to the About section of the MiSeq i100 Control Software to determine the control software version installed on the instrument.
Upgrade the Control Software to version ≥ 1.1 and restart the run setup.
MiSeq i100 Series Upgrade Software and release notes can be found in the MiSeq i100 Series Support Center: https://support.illumina.com/sequencing/sequencing_instruments/miseq-i100-plus/downloads.html
For any feedback or questions regarding this article (Illumina Knowledge Article #10125), contact Illumina Technical Support [email protected].
Last updated
Was this helpful?
