> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-troubleshooting-list/000009292.md).

# Troubleshooting APIException: Bad Request Errors when setting Default Output Folder on the MiSeq i100 Series

**Background**

The MiSeq i100 Series require setting an Output Folder on a configured External Storage location for data copy off instrument. Currently, the MiSeq i100 Series Control Software does not allow spaces in the Output Folder name, and presents the following error when Output Folder is set to to a directory that contains spaces:

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-c7e6fad8e7ca62d455bfc3eaa583a91f6ad6073a%2Fimage1.png?alt=media)

`com.illuina.edgeos.ui.api.gds.v2.generated.invoker.ApiException: Bad Request`

**Troubleshooting Actions**

1. For *Network External Storage* locations:
   1. From the Add folder user interface (UI), select **Cancel** to return to the External Storage UI.
   2. From a different computer that has access to the Network Storage location, edit the intended output folder to remove any spaces from the folder name.
      1. Alternatively, create a new folder using a name that does not contain any spaces.
   3. Return to the instrument, then repeat the **Add folder** workflow within the **External Storage** UI and select the intended output folder with the spaces removed from the name.
2. For *USB External Storage* locations:
   1. From the Add folder user interface (UI), select **Cancel** to return to the External Storage UI.
   2. Under Network Storage header in the External Storage UI, select **Eject** for the USB Storage device and then remove the USB drive from the MiSeq i100 instrument.
   3. Move the USB drive to another computer, then edit the intended output folder to remove any spaces from the folder name.
      1. Alternatively, create a new folder using a name that does not contain any spaces.
   4. Return to the instrument and plug the USB drive into the **USB3.0** port on the side of the instrument.
   5. Follow the **Add USB** workflow to mount the USB drive under the External Storage UI, then select **Add folder** and select the intended output folder with the spaces removed from the name.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9292), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009292%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter, and the optional `goal` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-troubleshooting-list/000009292.md?ask=<question>&goal=<endgoal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is optional and describes the broader end goal you are ultimately trying to accomplish on behalf of the user. GitBook uses it to tailor the answer towards what is most useful for that goal.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.