> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-reference_material-list/000010158.md).

# How to perform the Factory Restore on the MiSeq i100 Series

**Background**

The Factory Restore option allows Administrator users to restore the instrument to the default factory settings if there is an irrecoverable error on the MiSeq i100 Series instrument.

**Important Information**

The Factory Restore option **PERMANENTLY ERASES** all data on the instrument including Sequencing Run data, and should only be performed at the explicit recommendation of Illumina Technical or Field Support teams. The DRAGEN BCL Convert application is the only Application that is retained on Factory Restore, all other DRAGEN Applications must be re-installed after completing the Factory Restore and First Time Setup.

The Factory Restore process takes about 90 minutes and it cannot be canceled after it is initiated.

**Resolution Actions**

If the Factory Restore option is recommended by either Illumina Technical Support or Field Support teams, see the following steps to complete the process:

1. Login to the **MiSeq i100 Series Control Software** using a local **Administrator** account.
2. Select the **menu icon** in the top-left corner, then select **Settings**.
3. Select **Factory Restore**.
4. Review the information presented on the **Factory Restore** page, and select **Restore** to proceed.
5. Once complete, the instrument will **restart** to complete the process. Do NOT power cycle or restart the instrument during the Factory Restore process.
6. Once the Control Software has initialized, complete **First Time Setup**.
   1. Refer to the[First Time Setup](https://support-docs.illumina.com/IN/MiSeqi100Series/Content/IN/MiSeqi100/FirstSetup_SW.htm)section in the MiSeq i100 Series Product Documentation for details.
7. If additional **DRAGEN Applications** are required, download and re-install desired **DRAGEN Applications** and associated **Reference Genomes**.
   1. Refer to the [Applications](https://support-docs.illumina.com/IN/MiSeqi100Series/Content/IN/MiSeqi100/SysApps.htm) section in the MiSeq i100 Series Product Documentation for details.
8. Re-install the **DRAGEN License** files:
   1. *Online Installation*
      1. Login to the MiSeq i100 Series Control Software using a local **Administrator** account.
      2. Select the **menu icon** in the top-left corner, then select **Settings**.
      3. Select **DRAGEN**, then select the **License** tab.
      4. Select **Update from license server**.
      5. Once installation is complete, confirm that the **Valid Until** and **Serial Number** fields have populated.
   2. *Offline Installation*
      1. See Knowledge Article [Offline DRAGEN License Installation on the MiSeq i100 Series](https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-reference_material-list/000009557.md) for details.

If there are any issues during the Factory Restore, First Time Setup, or DRAGEN License and Application installation process, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                         |
| :---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #10158), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000010158%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-reference_material-list/000010158.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.